Ot sterilization chart method

Reading File
Finding Sources
Finding Sources
Searching PubMed

sterilization AND operating room

Searching the Web

operating theatre sterilization log chart autoclave monitoring chemical biological indicators guidance

OT sterilization chart method

An OT sterilization chart is a record of each sterilizer cycle. It documents that the load reached the required parameters and passed mechanical, chemical, and biological monitoring.
Use a separate entry for every load.
DateSterilizer IDLoad/cycle no.ContentsMethod/cycleTimeTemperaturePressureChemical indicatorBiological indicatorOperator/signatureRelease/remarks
22 Oct 2026Autoclave 1026Major OT instrument setsSteam, pre-vacuum4 min exposure132-135°CAs per machine cycleInternal and external CI passedWeekly BI pending / negativeName/signReleased
22 Oct 2026Autoclave 1027Linen packsSteam, gravity displacement30 min exposure121°CAs per machine cyclePassedN/AName/signReleased

How to complete it

  1. Before the cycle
    • Check cleaning, drying, correct packing, labeling, load arrangement, water/steam supply, and machine readiness.
    • Record the load number and contents.
  2. During or after the cycle: record mechanical parameters
    • For steam sterilization, document the cycle’s time, temperature, and pressure from the printout/display.
    • Ensure these meet the sterilizer and pack manufacturer’s validated instructions. Do not rely on one universal time-temperature setting for every item.
  3. Check chemical indicators for every pack/load
    • Use an external indicator on every pack, unless an internal indicator is visible.
    • Place an internal chemical indicator in each package or tray.
    • Record “pass” only if the indicator reaches its stated endpoint.
  4. Biological indicator or spore test
    • Use at least weekly, and preferably according to the hospital infection-control policy and manufacturer instructions.
    • For implant loads, use a biological indicator and quarantine the implant until the result is negative whenever feasible.
    • Record BI lot number, date, incubation/result, and the control result.
  5. Release or reject the load
    • Release only when mechanical readings and chemical indicators are satisfactory.
    • If parameters or indicators fail, mark the load NOT RELEASED, quarantine it, investigate the sterilizer/packing/loading issue, and reprocess instruments after correction.
    • A positive biological indicator requires the department’s recall and corrective-action protocol.

Daily Bowie-Dick test chart

For a pre-vacuum steam sterilizer, perform and document a daily Bowie-Dick air-removal test before processing routine loads.
DateSterilizer IDTest pack/lot no.ResultOperatorCorrective action if failed
22 Oct 2026Autoclave 1BD-118PassName/signNil
A failed Bowie-Dick test suggests inadequate air removal or a vacuum leak. Do not use the sterilizer for routine loads until evaluated and cleared.

Key principle

A chemical tape changing color does not itself prove sterility. Monitoring should combine:
  • Mechanical: time, temperature, pressure
  • Chemical: external and internal indicators
  • Biological: spore testing
This is consistent with CDC sterilizing-practice guidance, which recommends routine use of all three types of monitoring. The CDC summary recommendations also advises mechanical and chemical monitoring for each load and at least weekly biological-indicator monitoring.