Anaesthesia machine check protocol

Reading File
Finding Sources
Searching the Web

anaesthesia machine check protocol checklist preoperative 2024

Finding Sources
Reading Web Page

anaesthesia machine check protocol checklist steps

Finding Sources
Reading Web Page

anaesthesia machine check protocol steps high pressure low pressure breathing circuit scavenging

I now have comprehensive information from PMC, APSF (ASA 2008 guidelines), and the AAGBI/RCOA 2012 guidelines. Let me compile the full, structured answer.

Anaesthesia Machine Check Protocol

The pre-anaesthesia machine check is a mandatory safety procedure performed before every operating theatre session. Its purpose is to confirm the machine can reliably deliver the intended gas mixture, maintain ventilation, and support the patient's life. No single check works for every machine - modern automated systems handle some steps, but cannot replace all manual verification steps. The total time should not exceed 10 minutes for a full start-of-day check and under 2 minutes between cases.

Why It Matters

Equipment faults that were absent at the start of a case can develop during anaesthesia (pipeline failure, circuit disconnection, electrical failure, incorrect configuration). A machine check does not guarantee failure-free performance throughout the case - periodic checks during long procedures are also required.

A. Start-of-Day (Full) Check

Perform this at the beginning of each operating session, after the machine is moved, after servicing, or after vaporizers are changed.

1. Preparation

  • Ensure master switch is off, pipelines disconnected, all cylinders closed, and any residual gases exhausted (pressure gauges read zero).
  • Confirm backup equipment is present and functional: self-inflating bag (Ambu), auxiliary O2 cylinder, functioning suction.
  • Confirm the machine is connected to mains (AC) power. Battery backup should have at least 30 minutes of charge.

2. Gas Supply - High Pressure System (Cylinders)

StepAction
Identify cylindersConfirm identity by colour coding and label
O2 cylinderOpen fully (3.5 turns anticlockwise). Must be at least half full (>7,000 kPa / 1,000 psi)
Confirm O2 flowOpen O2 flow control valve - flowmeter should register flow
Cross-checkOpen N2O flow control valve - N2O flowmeter should not register flow when only O2 cylinder is open (confirms no cross-connection)
Close after checkClose O2 cylinder, wait for pressure to drop to zero
N2O cylinderCheck similarly; cylinder should be adequately filled
CO2 cylindersMust not be present on the machine. Fit blanking plug to any empty yoke
Nitrous interlockConfirm O2/N2O ratio protection device (hypoxic guard) is functional

3. Gas Supply - Pipeline System (Intermediate Pressure)

  • Reconnect pipeline supplies.
  • Check all pipeline pressure gauges read 400-500 kPa (approximately 58-73 psi or ≥50 psig).
  • Confirm O2 pipeline connects to the O2 inlet only (anti-cross-connection through Schrader probes or NIST fittings - these are non-interchangeable by design, but visual confirmation is still required).
  • Perform pipeline tug test to confirm secure connection.

4. Low Pressure System (Flowmeters and Vaporizers)

  • Turn master switch on.
  • If the machine has minimum mandatory flows, confirm O2 flowmeter registers 50-200 mL/min with the O2 flow control knob turned off.
  • Test all flowmeters: look for bobbin/ball damage, sticky floats, and correct full range of flow.
  • Check vaporizers are adequately filled; confirm filler ports are tightly closed.
  • Check vaporizer interlock (only one vaporizer can be on at a time).
  • Perform low-pressure system leak test (negative pressure / syringe test):
    • With the vaporizer off and common gas outlet occluded, compress a syringe attached to the outlet - no refill should occur (leak <100 mL/min).
    • Repeat for each vaporizer individually (turn each on during the test, then off).
    • Leaks as low as 100 mL/min can critically reduce volatile agent concentration and risk intraoperative awareness.

5. Oxygen Monitoring and Fail-Safe

  • Calibrate the O2 analyser (or verify calibration):
    • Reads ~21% in room air.
    • Reads >95% in 100% O2 flow.
  • Check the low O2 alarm activates at the set threshold (typically ≤25-30%).
  • Confirm the oxygen failure alarm (Ritchie whistle / audible alarm) sounds when O2 supply pressure drops.
  • Confirm the hypoxic guard (proportioning system) prevents delivery of <21% O2.

6. Breathing System (Circle System or T-piece)

  • Inspect the circuit for correct configuration and assembly: no missing components, all connections tight.
  • Check CO2 absorber (soda lime/Amsorb):
    • Confirm it is not exhausted (colour change indicator, adequate volume).
    • No dry or overheated absorbent.
  • Perform pressure/leak test of the breathing circuit:
    1. Occlude the patient port.
    2. Fill the system to 30 cmH2O with the O2 flush.
    3. Close the APL (pop-off) valve.
    4. No pressure drop over 10 seconds = no significant leak.
  • Check APL (adjustable pressure-limiting) valve opens and closes correctly.
  • Check unidirectional valves (inspiratory and expiratory) are present and moving freely.
  • Confirm gas flows correctly through both inspiratory and expiratory limbs.
  • Check filters and HME connectors are fitted correctly.

7. Ventilator

  • Configure ventilator for intended use (volume control, pressure control as needed).
  • Ensure ventilator tubing is securely attached.
  • Set initial controls and confirm adequate inspiratory pressure is generated.
  • Confirm alarms are active and correctly set: high pressure, low pressure (disconnect alarm), minute volume.
  • Check pressure relief valve functions at the set pressure.
  • Confirm bellows or piston moves correctly and returns to baseline.

8. Scavenging System

  • Confirm the anaesthetic gas scavenging system (AGSS) is switched on and functioning.
  • Ensure tubing is attached to the correct exhaust port of the breathing system, ventilator, or workstation.
  • Check for no obstruction in the scavenging pathway (a blocked scavenging system can cause dangerously raised airway pressures).

9. Monitoring Equipment

MonitorCheck
SpO2Apply to self - should read >96%
Capnograph (EtCO2)Exhale into sampling port - should register a CO2 waveform
NIBPCycle on arm - confirm reading
Airway pressure gaugeConfirm visible and functioning
Gas analyserConfirm sampling line connected and patent
All alarmsActivate and confirm audible/visual alerts work; set to patient-appropriate thresholds

10. Suction

  • Confirm suction is available, tubing is connected, and generates adequate negative pressure.
  • Yankauer sucker should be at hand.

11. Ancillary Equipment and Drugs

  • Check airway equipment for the planned case: laryngoscope (blade light working, spare batteries), ETT (cuff inflates/deflates), LMA, bougie, CICO (can't intubate, can't oxygenate) rescue equipment.
  • Confirm drugs drawn up or available: induction agent, muscle relaxant, reversal agents, vasopressors, analgesics.
  • Verify sufficient CO2 absorbent and volatile agent in the vaporizer.

12. Documentation

  • Record completion of the check with the date, time, and name of the checking anaesthetist on the machine log.
  • This is a medicolegal requirement.

B. Between-Case Check (Start of Each Patient)

These items must be re-verified before every new patient:
#ItemResponsible
1Suction adequateProvider + Tech
2Monitors available and alarms setProvider or Tech
3Vaporizers filled, filler ports closedProvider
4CO2 absorbent not exhaustedProvider or Tech
5Breathing system pressure/leak testProvider + Tech
6Gas flows correctly through circuit (inspiration and expiration)Provider + Tech
7Document completionProvider + Tech
Whenever a breathing system is changed during a list (especially in paediatric cases), its integrity and correct configuration must be re-confirmed before use.

C. During a Long Case

  • Periodically check for exhaustion of volatile agent in the vaporizer.
  • If an O2 cylinder was used during the case, confirm it is at least half full before the next patient or replace it.
  • On handover between anaesthetists mid-case, formal machine check status must be communicated.

D. After a Critical Incident

Even if the machine was checked before the case, perform an immediate brief check after any critical intraoperative incident - the incident itself may have been caused by an equipment fault that developed during the case.

Key Reference Values Summary

ParameterAcceptable Value
O2 cylinder pressure>7,000 kPa (>1,000 psi, at least half full)
Pipeline pressure400-500 kPa (≥50 psig)
O2 analyser in air~21%
O2 analyser in 100% O2>95%
Low O2 alarm threshold≤25-30%
Breathing circuit pressure testHold 30 cmH2O for 10 seconds, no drop
Machine mandatory O2 flow (if applicable)50-200 mL/min
Battery backup minimum30 minutes

Sources

Note: Always supplement this protocol with your institution's specific checklist and the manufacturer's instructions for your machine model, as automated self-checks vary significantly between systems and do not replace manual verification of all items.
This is a shared conversation. Sign in to Orris to start your own chat.