Multisection tomography in detail
| Era | Technology | Key Limitation |
|---|---|---|
| Pre-1990 | Sequential (step-and-shoot) single-slice CT | Inter-slice gaps; patient must hold breath for each slice |
| ~1990 | Spiral (helical) single-slice CT | Volumetric acquisition possible but slow; thick slices |
| 1998 | 4-slice MDCT introduced by multiple manufacturers | Began the MDCT era |
| 2000s | 16- → 64-slice MDCT | Near-isotropic voxels; single breath-hold chest |
| Present | 128-, 256-, 320-detector-row systems; dual-source | Temporal resolution <100 ms; whole-organ coverage in one rotation |
"In 1998 several CT manufacturers introduced multidetector systems, which provided considerable improvement in acquisition speed, coverage, and temporal and spatial resolution." — Grainger & Allison's Diagnostic Radiology
| Technology | Mechanism | Benefit |
|---|---|---|
| Dual-source CT | Two x-ray tubes at 90° to each other | Temporal resolution ~75 ms; dual-energy capability |
| Flying focal spot | Alternates focal spot position rapidly | Converts 128-row → virtual 256-row array; improved z-resolution |
| Wide-area detector (320-row) | 16 cm z-axis coverage per rotation | Entire heart or organ imaged in a single rotation |

| Algorithm type | Effect | Clinical use |
|---|---|---|
| Low spatial frequency (smooth) | Reduces noise; lower spatial resolution | Soft tissue, vascular structures, mediastinum |
| High spatial frequency (sharp/lung) | Enhances fine detail; increases noise | Lung parenchyma, airways, bone, HRCT |
| Technique | Method | Key Application |
|---|---|---|
| MPR / CMPR (multiplanar/curved MPR) | 1-voxel-thick tomographic sections in any plane | Pulmonary emboli, airways evaluation |
| MIP (maximum intensity projection) | Only highest-attenuation voxels rendered | Vascular imaging, micronodular disease distribution |
| MinIP (minimum intensity projection) | Only lowest-attenuation voxels rendered | Emphysema, air trapping (augments air-trapping conspicuity) |
| Shaded surface display (SSD) | Threshold-based surface rendering | Large airway/vessel overview |
| Volume rendering (VR) | All voxels, opacity assigned by HU | Surgical planning, complex anatomy |
| Virtual bronchoscopy | Fly-through of airway lumen | Airway stenosis evaluation |
| Mode | Mechanism | Radiation dose | Use |
|---|---|---|---|
| Prospective gating | Scan triggered at fixed interval after R-wave; pauses between beats | Low (<1 mSv achievable) | Calcium scoring, coronary CTA at stable HR |
| Retrospective gating | Continuous acquisition throughout cardiac cycle; retrospective phase selection (0–90% R-R at 10% intervals) | High (~10× prospective) | Functional assessment (ejection fraction, wall motion) |
| Category | Strategies |
|---|---|
| Tube current | AEC; weight/size-based modulation; ECG gating; low-dose protocols (40–70 mAs for chest) |
| Tube voltage | Reduced kVp (especially in thin patients or paediatrics) |
| Beam shaping | Bowtie filters |
| Acquisition | Higher pitch; shorter z-coverage; faster rotation; wider collimation |
| Reconstruction | Iterative reconstruction (especially model-based); high-sensitivity detectors |
| Protocol | Patient shielding; restrict FOV to area of interest |
No I am talking about thisHere are **complete, exam-oriented, concept + mechanism based notes on Multisection (Multilayer) Tomography**, strictly built from your provided pages and expanded for clarity without losing any detail. --- # 🌟 MULTISECTION TOMOGRAPHY ## 📌 1. Definition **Multisection (or multilayer) radiography** is a type of tomography in which **multiple body layers are imaged simultaneously**, with **each layer recorded on a separate film** during a **single exposure**. 👉 Key idea: Instead of taking one slice at a time → **many slices are obtained at once** --- ## ⚙️ 2. Principle (Core Concept) ### 🔑 Basic Tomography Principle: * X-ray tube (T) and film (F) move in **opposite directions** around a **fulcrum (pivot point)**. * The **plane passing through the fulcrum (Y)** remains **sharp**. * Structures above and below → blurred. --- ### 🔑 Multisection Extension: In multisection tomography: * Multiple films are placed **at different vertical levels**. * Each film has its **own effective fulcrum**. 👉 So: * Film at level F₁ → records plane at fulcrum Y * Film at level F₂ → records plane at fulcrum Y₂ * Additional films → create additional focal planes --- ### 🧠 Conceptual Diagram ``` Tube movement T1 → T2 \ / \ / \ / \ / Y ← Fulcrum for upper film (F1) / \ / \ / \ Y2 ← Fulcrum for lower film (F2) / \ F1 F2 ← Films at different levels ``` --- ### 🔑 Important Principle Statement (from your text): > “The layer recorded on a film is at the level of the fulcrum of the lever system, provided the film position coincides with the film-moving point on the lever.” ✔️ Films above fulcrum → record higher planes ✔️ Films below fulcrum → record lower planes --- ## ⚙️ 3. Technique ### 🧪 Step-by-step Working: 1. **Arrangement of Films** * Multiple films stacked vertically * Separated by spacers 2. **Tube-Film Motion** * X-ray tube and film move in opposite directions * Rotation occurs around fulcrum(s) 3. **Simultaneous Exposure** * Single exposure produces multiple images 4. **Plane Formation** * Each film corresponds to a different focal plane * Achieved due to different effective fulcrum levels --- ### 🔬 Important Observations: ✔ All radiographs are taken **at exactly the same moment** → Same respiratory phase → No motion mismatch ✔ Useful for **rapid transient phenomena** → Example: angiography (vascular filling) --- ## ⭐ 4. Advantages 1. **Reduced radiation dose** * Multiple images in single exposure 2. **Same physiological phase imaging** * No breathing mismatch 3. **Best for transient processes** * e.g., angiography 4. **Time saving** * For patient and department 5. **Reduced tube loading** * Less repeated exposures --- ## ⚠️ 5. Limitation ❗ Sequential tomography (one section at a time) gives: → **Better detail (higher resolution)** 👉 So: Multisection = faster Sequential = more detailed --- # 🧰 6. EQUIPMENTS (VERY IMPORTANT) ## 🟦 A. Multisection Cassette ### 📦 Structure: * A **metal box (cassette)** containing multiple films * Depth: up to **7.5 cm (3 inches)** * Holds **3, 4, 5 or 7 films** --- ### 📏 Film Spacing: * Films separated by fixed distance * Example: **5 mm spacing for 4 films** --- ### ⚙️ Placement: * Shallow cassette → fits in standard bucky tray * Deep cassette → requires **special support tray** --- ### 🧠 Function: * Maintains fixed geometry * Allows simultaneous imaging of multiple layers --- ## 🟦 B. Intensifying Screens & Separators ### 📚 “Book Arrangement” Concept: * Screens + spacers assembled like a **book** * Films inserted between “pages” --- ### ✅ Advantages: ✔ Prevents wrong film placement ✔ Ensures proper alignment ✔ Easy handling --- ### ⚠️ Density Problem (VERY IMPORTANT CONCEPT) As X-ray beam passes through cassette: * It gets **progressively attenuated** * Lower films receive **less radiation** 👉 Result: * Upper film → more exposure * Lower film → less exposure --- ### 🔧 Solution: * Use **different speed intensifying screens** ✔ Lower films → **faster screens** ✔ Upper films → **slower screens** 👉 To maintain **uniform density** --- ### 🧠 Key Statement: > “The last intensifying screens must be greater in speed than those nearer to the X-ray tube.” --- ## 🟦 C. Separators ### 📌 Material: * Usually **plastic foam** ### 📏 Thickness: * Typically **5 mm or 10 mm** * Can be reduced to **1 mm** in special cases --- ### ⚠️ Requirements: ✔ Should NOT absorb X-rays significantly ✔ Must be **artifact-free** --- ## ⚠️ 7. Exposure Considerations ### 📉 System Speed: * Multisection system is **photographically slower** 👉 Why? * Multiple layers + attenuation * Complex screen-film interaction --- ### 📊 Exposure Adjustment: * kVp increase: **~8–19 kVp** * mAs increase: **~1.6 to 2.5 times** --- ### 🧠 Important: * No fixed formula → depends on setup * Often determined by: 👉 **Trial and error** 👉 **Step wedge experiments** --- ### ⚠️ Practical Note: * First-time users should assume: 👉 System is **4–5 times slower** than standard cassette --- ## ⚠️ 8. Practical Errors ❗ Wrong orientation of “book” cassette: → Leads to **radiographic chaos** 👉 Always ensure: * Correct side facing X-ray tube * Proper identification marking --- # 🎯 FINAL SUMMARY (Quick Revision) | Feature | Multisection Tomography | | ---------- | ------------------------------ | | Imaging | Multiple layers simultaneously | | Exposure | Single | | Fulcrum | Multiple (one per film level) | | Advantage | Time + dose saving | | Limitation | Less detail than sequential | | Cassette | Multi-film stacked system | | Problem | Unequal density | | Solution | Variable speed screens | --Got it — you want **fully detailed, point-wise notes** (no skipping, no compression). Here’s a **structured, exam-ready breakdown** of *Multisection Tomography* strictly based on your pages + explained clearly. --- # 🌟 MULTISECTION (MULTILAYER) TOMOGRAPHY — DETAILED POINTS --- # 📌 1. Definition 1. Multisection (multilayer) radiography is a form of tomography. 2. It involves imaging of **multiple body layers simultaneously**. 3. Each layer is recorded on a **separate film**. 4. All films are exposed **in a single exposure**. 5. It is designed to overcome limitations of single-layer tomography. --- # ⚙️ 2. Principle ### 🔑 Basic Tomography Principle 1. X-ray tube (T) and film (F) are connected by a **mechanical lever system**. 2. They move in **opposite directions** during exposure. 3. Movement occurs around a fixed point called the **fulcrum (Y)**. 4. The plane passing through the fulcrum: * Remains **stationary relative to film** * Appears **sharp (in focus)** 5. Structures above and below: * Move relative to film * Become **blurred** --- ### 🔑 Multisection Principle (Extension) 6. In multisection tomography, **multiple films** are used. 7. Films are placed at **different vertical levels**. 8. Each film has a **different effective fulcrum**. 9. Example: * Film F₁ → records plane at fulcrum Y * Film F₂ → records plane at fulcrum Y₂ (lower level) 10. Each additional film creates its **own focal plane**. --- ### 🔑 Key Principle Statement 11. The layer recorded on a film corresponds to the **level of the fulcrum**. 12. This is valid only if: * Film position coincides with **film-moving point on lever** 13. Films above this point: * Record **higher planes** 14. Films below: * Record **lower planes** --- ### 🔬 Important Concept 15. Multiple fulcrum levels are **created geometrically**, not physically shifted. 16. Each film “selects” a different anatomical layer due to position difference. --- # ⚙️ 3. Technique ### 🧪 Film Arrangement 1. Multiple films are arranged: * One above the other 2. They are separated by: * **Spacing material (separators)** --- ### 🧪 Movement 3. X-ray tube moves in one direction. 4. Films move in the opposite direction. 5. Movement is synchronized via **lever system**. --- ### 🧪 Exposure 6. Only **one exposure** is given. 7. All films are exposed **simultaneously**. --- ### 🧪 Image Formation 8. Each film records a **different tomographic layer**. 9. The layer corresponds to: * Its **relative position in cassette** 10. Blurring occurs for: * Structures not in that plane --- ### 🧪 Physiological Advantage 11. All images are taken: * At the **same moment** 12. Therefore: * Same respiratory phase * Same cardiac phase (if relevant) --- ### 🧪 Special Use 13. Useful in **rapid transient processes** * Example: angiography (vascular filling) --- # ⭐ 4. Advantages 1. **Reduced radiation dose** * Multiple images from single exposure 2. **Same time imaging** * No variation due to breathing/motion 3. **Captures transient phenomena** * e.g., contrast flow 4. **Time-saving** * For patient and radiology department 5. **Reduced X-ray tube load** * Fewer exposures needed --- # ⚠️ 5. Disadvantage 1. Sequential tomography (single layer at a time): * Provides **better detail** 2. Multisection: * Slightly **lower image sharpness** --- # 🧰 6. Equipment --- ## 🟦 A. Multisection Cassette ### 📦 Structure 1. It is a **metal box**. 2. Designed to hold **multiple films**. 3. Depth: * Up to **7.5 cm (3 inches)** --- ### 📦 Capacity 4. Available models: * 3 films * 4 films * 5 films * 7 films --- ### 📦 Film Spacing 5. Films are separated by fixed distances. 6. Example: * 4 films → spaced **5 mm apart** --- ### 📦 Placement 7. Shallow cassette: * Fits into **standard bucky tray** 8. Deep cassette: * Requires **special tray or drawer** --- ### 📦 Function 9. Maintains: * Alignment * Proper geometry 10. Allows simultaneous imaging of multiple layers --- ## 🟦 B. Intensifying Screens & Separators --- ### 📚 Book Arrangement 1. Screens and separators are fixed together. 2. Arranged like a **book**. 3. Films inserted between layers (like pages). --- ### ✅ Advantages 4. Prevents incorrect film placement. 5. Ensures correct alignment with screens. --- ### ⚠️ Unequal Density Problem 6. X-ray beam passes through multiple layers. 7. It is **progressively attenuated**. 8. Lower films receive: * Less radiation 9. Upper films receive: * More radiation --- ### 🔧 Solution 10. Use **different speed intensifying screens**: * Upper films → slower screens * Lower films → faster screens 11. Ensures: * Uniform density across all films --- ### 🧠 Key Concept 12. Screens farther from X-ray tube must be: * **Higher speed** --- ## 🟦 C. Separators --- ### 📌 Material 1. Usually made of: * **Plastic foam** --- ### 📏 Thickness 2. Common thickness: * 5 mm * 10 mm 3. Special cases: * As low as **1 mm separation** --- ### ⚠️ Requirements 4. Must NOT: * Absorb significant X-rays 5. Must: * Be free from artifacts --- # ⚠️ 7. Exposure Factors --- ### 📉 System Speed 1. Multisection system is: * **Photographically slower** --- ### 📉 Reason 2. Due to: * Multiple layers * X-ray attenuation * Screen-film complexity --- ### 📊 Exposure Adjustment 3. kVp increase: * Approximately **8–19 kVp** 4. mAs increase: * Approximately **1.6 to 2.5 times** --- ### ⚠️ Practical Issue 5. No fixed exposure formula. 6. Depends on: * Cassette type * Film-screen combination --- ### 🧪 Best Method 7. Use: * **Trial and error** * **Step wedge experiments** --- ### ⚠️ Important Note 8. System may be: * **4–5 times slower** than standard cassette --- # ⚠️ 8. Practical Considerations --- ### ❗ Orientation Error 1. Cassette has correct and wrong side. 2. If placed incorrectly: * Produces **radiographic errors (chaos)** --- ### ❗ Identification 3. Proper marking is provided. 4. Must ensure: * Correct side faces X-ray tube --- # 🎯 FINAL QUICK POINTS 1. Multiple films → multiple layers 2. Single exposure → reduced dose 3. Different film levels → different fulcrum planes 4. Unequal exposure → corrected by screen speed 5. System slower → increase kVp & mAs 6. Cassette design → key role 7. Sequential tomography → better detail --- If you want next step (highly recommended for exams):
"The layer recorded on a film is at the level of the fulcrum of the lever system, provided the film position coincides with the film-moving point on the lever."
| Advantage | Explanation |
|---|---|
| Reduced radiation dose | Multiple slices from one exposure vs. multiple separate exposures in sequential tomography |
| Same physiological moment | All slices represent identical respiratory phase, cardiac phase, and vascular filling phase — no mismatch |
| Captures transient phenomena | Ideal for angiography — all slices show the same moment of contrast filling |
| Time efficiency | Saves examination time for patient and department |
| Reduced X-ray tube loading | Fewer total exposures → less tube wear and heat accumulation |
| Feature | Detail |
|---|---|
| Construction | Metal box designed to hold multiple films stacked vertically |
| Depth | Up to 7.5 cm (3 inches) |
| Capacity | 3, 4, 5, or 7 films |
| Film spacing | Fixed; example: 5 mm between 4 films |
| Shallow cassette | Fits standard bucky tray |
| Deep cassette | Requires special support tray or drawer |
| Function | Maintains fixed geometry and alignment; enables simultaneous multi-layer imaging |
X-ray beam enters from above
↓
Film 1 (top) ← receives MOST radiation
↓ [attenuated]
Film 2 ← receives less
↓ [further attenuated]
Film 3 ← receives even less
↓ [further attenuated]
Film 4 (bottom) ← receives LEAST radiation
| Film position | Screen speed required |
|---|---|
| Top (nearest X-ray tube) | Slow screens (require more radiation to produce same light output) |
| Bottom (farthest from tube) | Fast screens (produce more light per unit of radiation) |
"The last intensifying screens must be greater in speed than those nearer to the X-ray tube."
| Property | Detail |
|---|---|
| Material | Plastic foam (preferred) |
| Standard thickness | 5 mm or 10 mm |
| Special thin separators | As low as 1 mm (for closely spaced planes) |
| X-ray absorption | Must be minimal — should not absorb significant radiation |
| Artefact requirement | Must be artefact-free — any structure in the foam would appear on the films |
| Parameter | Adjustment |
|---|---|
| kVp | Increase by approximately 8–19 kVp |
| mAs | Increase by approximately 1.6–2.5 times |
| Error | Consequence | Prevention |
|---|---|---|
| Cassette inverted (wrong side facing tube) | Film order reversed; incorrect planes labelled; radiographic "chaos" | Always check orientation markings before placement |
| Incorrect screen-film matching | Density variation across films; diagnostic failure | Follow manufacturer's loading guide; use book arrangement |
| Insufficient exposure increase | Lower films underexposed | Apply correct kVp/mAs uplifts; verify with step wedge |
| Separator with artefacts | Artefact lines or densities on films | Inspect separators before use; replace damaged foam |
| Feature | Multisection | Sequential |
|---|---|---|
| Exposures per session | One | Multiple (one per slice) |
| Number of slices | Multiple simultaneously | One at a time |
| Radiation dose | Lower | Higher |
| Image detail/resolution | Lower | Higher |
| Physiological phase consistency | Perfect (same instant) | Variable (different breath-holds) |
| Best for | Transient phenomena, angiography, dose reduction | When fine detail is the priority |
| Tube loading | Less | More |
| Time | Faster | Slower |
| Concept | Key Point |
|---|---|
| Core mechanism | Multiple films at different heights → different effective fulcra → different focal planes |
| Single exposure | All films exposed simultaneously → identical physiological moment |
| Governing principle | Focal plane = level of fulcrum for that film's position on the lever |
| Films above lever point | Record higher planes |
| Films below lever point | Record lower planes |
| Cassette depth | Up to 7.5 cm; holds 3–7 films |
| Film spacing example | 4 films at 5 mm intervals |
| Separator material | Plastic foam; must be radiolucent and artefact-free |
| Density problem | Progressive beam attenuation → lower films underexposed |
| Solution | Faster screens for lower films; slower screens for upper films |
| Exposure increase | kVp +8–19; mAs ×1.6–2.5 |
| Starting assumption | System 4–5× slower than standard cassette |
| Key disadvantage | Less image detail than sequential tomography |
| Key advantage | Dose saving + physiological phase consistency |
Give me notes on scan converter or scan converter memory and also define difference in two , image formation, image storage, display in ultrasound in detail and steps how it done and give me notes as u give to a msc radiology student
| Feature | Detail |
|---|---|
| Storage medium | Electrical charge on a physical storage tube |
| Data format | Analogue (continuous, not discrete) |
| Coordinate conversion | Achieved geometrically by controlling electron beam deflection |
| Grey scale | Limited — typically 8–16 grey levels |
| Noise | High — charge distribution is unstable; image degrades over time |
| Flicker | Present — the stored image "fades" as charge dissipates |
| Image freeze | Possible but image quality deteriorated rapidly |
| Portability | Low — bulky, fragile glass tube |
| Post-processing | None — image was fixed once written |
| Component | Function |
|---|---|
| Analog-to-Digital Converter (ADC) | Converts analogue echo voltage signals → binary numbers |
| Write Memory (Frame Buffer) | Stores the incoming scan-line data as pixel values |
| Scan Conversion Algorithm | Maps polar coordinate scan-line data → rectangular pixel matrix |
| Interpolation Circuit | Fills in gaps in the rectangular matrix between scan lines |
| Read Memory | Reads stored matrix data and sends to display |
| Digital-to-Analog Converter (DAC) | Converts digital pixel values → analogue video signal for monitor |
| Feature | Analog Scan Converter | Digital Scan Converter |
|---|---|---|
| Storage medium | Electrical charge on CRT storage tube | Binary numbers in RAM |
| Data type | Continuous analogue | Discrete digital |
| Grey levels | 8–16 | 256 (8-bit) or more |
| Image stability | Unstable — charge fades, flickers | Perfectly stable — no degradation |
| Image freeze | Possible but degrades rapidly | Perfect freeze — indefinite |
| Noise | High | Low |
| Post-processing | None | Full — zoom, edge enhancement, filtering, measurements |
| Geometric accuracy | Limited by beam deflection mechanics | High — mathematical coordinate conversion |
| Image storage | Temporary — charge lost on power off | Permanent — can archive to disk/PACS |
| Quantitative analysis | Not possible | Full — distance, area, volume, Doppler measurements |
| Portability | Bulky, fragile | Compact, robust |
| Current use | Obsolete | Standard in all modern machines |
| Post-processing Function | Effect |
|---|---|
| Grey-scale mapping (post-processing curves) | Remaps digital values to different grey-scale output — adjust contrast of stored image |
| Edge enhancement | Sharpens boundaries between structures |
| Smoothing/averaging | Reduces speckle noise |
| Zoom (write vs. read zoom) | Write zoom: acquires data from smaller region → higher resolution. Read zoom: magnifies stored data → no added resolution |
| Frame averaging (persistence) | Averages 2–4 consecutive frames → reduces noise but reduces temporal resolution |
| Measurements | Electronic callipers read pixel coordinates → calculate distance, area, volume |
| Colour assignment | Assigns false colour maps to grey-scale values |
| Method | Format | Notes |
|---|---|---|
| DICOM (Digital Imaging and Communications in Medicine) | .dcm files | Universal standard; preserves raw pixel data + patient metadata |
| PACS (Picture Archiving and Communication System) | Network server storage | Central archive; accessible from multiple workstations |
| Local hard disk | Proprietary or DICOM | On-machine storage |
| USB / optical media | JPEG, MPEG, AVI, or DICOM | For portable review |
| Cine loops | MPEG/DICOM video | Essential for cardiac imaging — captures motion |
TRANSDUCER
│
▼ Electrical pulse → Ultrasound pulse
TISSUE INTERACTION
│
▼ Reflected echo → Electrical signal
PRE-AMPLIFIER
│
▼ Amplified RF signal
TIME-GAIN COMPENSATION (TGC)
│
▼ Depth-corrected signal
FILTERING → RECTIFICATION → ENVELOPE DETECTION
│
▼ Amplitude profile per scan line
COMPRESSION (Dynamic range reduction)
│
▼ Compressed analogue signal
ADC (Analog-to-Digital Converter)
│
▼ Digital values per scan line (polar coordinates)
DIGITAL SCAN CONVERTER
│
├─ Scan conversion (polar → rectangular)
├─ Interpolation (fill matrix gaps)
▼
FRAME BUFFER (Digital Image Matrix: 512×512 or 1024×1024 pixels)
│
├─ Post-processing (grey map, zoom, averaging, measurements)
▼
DAC (Digital-to-Analog Converter)
│
▼
DISPLAY MONITOR (Grey-scale real-time image)
│
▼
ARCHIVING (PACS / DICOM / Cine loop)
| Bit depth | Grey levels | Clinical implication |
|---|---|---|
| 6-bit | 64 | Older systems; limited tissue differentiation |
| 8-bit | 256 | Current standard; adequate for most clinical use |
| 10-bit | 1024 | High-end systems; better differentiation of low-contrast structures |
| 12-bit | 4096 | Research/premium systems |
Frame rate = PRF ÷ (number of scan lines per frame)
PRF(max) = 1540 m/s ÷ (2 × imaging depth in metres)
| Concept | Key Point |
|---|---|
| Why scan converter is needed | Polar scan-line data ≠ rectangular monitor pixel matrix |
| Analog scan converter | Charge on storage tube; 8–16 grey levels; fades, flickers; obsolete |
| Digital scan converter | Binary values in RAM; 256 grey levels; stable, post-processable; current standard |
| ADC | Converts analogue echo → digital number |
| Frame buffer | Digital RAM matrix storing one complete image frame |
| Scan conversion | Mathematical mapping of (r,θ) → (x,y) using x = r·sinθ, y = r·cosθ |
| Interpolation | Fills pixel gaps between scan lines |
| TGC | Corrects for depth-dependent attenuation |
| Compression | Reduces wide dynamic range of echoes to displayable grey-scale range |
| Post-processing | Manipulates stored digital image — grey map, zoom, averaging |
| DICOM | Universal standard for ultrasound image storage and communication |
| Frame rate | PRF ÷ number of scan lines; limited by depth |
| 8-bit grey scale | 256 grey levels; current standard |