CONDUCTING LOW‑FLOW ANAESTHESIA
Premedication, preoxygenation and induction of sleep
are performed according to the usual practice.
Initiation of low‑flow anaesthesia
The objective is to achieve an alveolar concentration
of the anaesthetic agent that is adequate for producing
surgical anaesthesia. There are different methods of
achieving this objective.
Use of high flows during initial phase
The time constant is reduced, bringing the circuit
concentration to the desired concentration rapidly.
Often, an FG flow of 10 L of the desired gas concentration
and 2 MAC agent concentration is used. By the end
of 3 min (i.e., 3 time constants), the circuit would be
brought to the desired concentration. This facilitates
better denitrogenation and rapid achievement of
desired concentration by counterbalancing the large
uptake encountered at the start of the anaesthesia.
Use of prefilled circuits
Here, we use a different circuit like Magill’s for
preoxygenation. Simultaneously, the circle system is
fitted with a test lung and the entire circuit is filled
with the gas mixture of the desired concentration.
After tracheal intubation, the patient is connected to
the circle system and rapid achievement of the desired
concentration in the circuit occurs.
Injection of volatile agent into the breathing circuit
The usual requirement of anaesthetic agent is
approximately 400–500 ml of vapour in the first
10 min (i.e., 40–50 ml/min). At 20°C, 1 ml liquid
halothane yields 226 ml of vapour and 1 ml isoflurane
yields 196 ml. About 2 ml of the liquid agent is
injected in small increments into the expiratory limb
of the circuit. The intermittent injections are often
made in 0.2–0.5 ml aliquots manually. Alternatively,
continuous infusion may be used with the added
advantage of doing away with the peaks and troughs
associated with intermittent injections. The accurate
dose requirement is given by the formula:
Priming dose (ml vapour) = Desired concentration ×
([FRC + circuit volume] + [cardiac output
×
blood gas
coefficient])
Maintenance of low‑flow anaesthesia
During this phase, we need to maintain steady‑state
concentration of the anaesthetic agents. Although the
oxygen uptake remains constant at 200–250 ml/min, uptake of anaesthetic agents including N2O will be
minimal. Therefore, the role of oxygen analyser to
maintain oxygen concentration of at least 30% at all
times is paramount. It is prudent to return the sampling
gas (usually drawn at the rate of 200 ml/min) back to
the circuit to boost the economy of FGF utilisation.
It should be noted that the actual dial setting in the
vapourisers often over‑estimates the actual output
since the plenum vapourisers under delivers the agent
at low flows. The achievement of the desired end tidal
agent concentration may be measured most accurately
using an agent analyser or by the haemodynamic
stability.
The Gothenburg technique[20] of conducting LFA is
depicted in Figure 1.
Termination of low flow anaesthesia
Because of long‑time constants, recovery is delayed
in LFA. However, switching over to high flows to
accelerate the wash‑out of anaesthetic agents or use
of activated charcoal to remove the potent vapours by
absorption can result in rapid recovery. Nitrous oxide
gets washed off while changing over to 100% oxygen.