Pns waterview postioning and tube distance
Waters view PNS paranasal sinuses positioning radiograph
PNS Waters view positioning tube distance radiography technique
https://radiologykey.com/radiological-techniques-protocols-a…
PNS Waters view occipitomental radiograph positioning
| Parameter | Detail |
|---|---|
| Posture | Patient erect, facing the image receptor (PA projection) |
| Chin | Raised/extended, resting chin against the bucky/detector |
| Midsagittal plane | Perpendicular to the receptor (no rotation) |
| Line alignment | Mentomandibular Line (MML) perpendicular to the receptor |
| Result | Orbitomeatal Line (OML) forms 37° angle with the image receptor |
| Head position | Slightly extended, looking slightly upward |
Key tip: The patient may need to "slouch forward" or do a slight "limbo" position to get the chin high enough. Check from above for any tilt or rotation.
| Parameter | Detail |
|---|---|
| Beam direction | Horizontal (perpendicular to the image receptor, 0° angulation) |
| Centering point | CR exits at the acanthion (base of the anterior nasal spine, between nose and upper lip) |
| SID (Source-to-Image Distance) | 100 cm (40 inches) |
| Grid | Yes (departmental variation applies) |
| Parameter | Value |
|---|---|
| kVp | 75-80 kVp |
| mAs | 20-25 mAs |
| Detector size | 24 cm x 30 cm |
| Orientation | Portrait |
| Feature | Waters' View | Caldwell's View |
|---|---|---|
| Primary target | Maxillary sinuses | Frontal sinuses |
| Tube angulation | 0° (horizontal beam) | 15° caudad |
| OML angle to film | 37° | ~15° |
| Petrous ridges | Below maxillary sinuses | Lower 1/3 of orbits |
| CR exits at | Acanthion | Nasion |