I now have excellent source material. Let me compile the comprehensive answer on all MRI artifacts from all sources gathered, including the images visible in the attachments.
MRI ARTIFACTS - Complete Reference (50 Marks)
PHYSICS FOUNDATION
MRI image formation relies on three core processes:
- Spin excitation - RF pulse at Larmor frequency (f = γ × B₀, where γ = 42.57 MHz/T for protons)
- Spatial encoding - gradient fields along frequency-encode (FE), phase-encode (PE), and slice-select directions
- Signal detection - raw data fills k-space; Fourier transform converts k-space to image
Any disruption to these processes produces an artifact. Understanding which step is disrupted explains both the appearance and the cure.
CATEGORY 1 - MOTION-RELATED ARTIFACTS
1. Motion Artifact (Ghosting)
Appearance: Blurring, ghosting, smearing across the image - typically in the phase-encode direction.
Mechanism:
- MRI acquires multiple phase-encoding lines sequentially over time (each TR acquires one k-space line in conventional spin echo).
- If tissue moves between phase-encoding steps, each k-space line is acquired from a slightly different tissue position.
- This produces phase inconsistency in k-space.
- Fourier reconstruction treats these phase differences as if they arise from different spatial frequencies, projecting "ghost" copies of moving structures displaced in the PE direction.
- Ghost spacing = FOV / (motion frequency × TR × number of PE steps).
Voluntary motion: patient movement
Involuntary motion: respiration, cardiac pulsation, bowel peristalsis, vascular flow
Reduction:
- Patient education: instruct to hold still, practice breath-holding
- Sedation or general anesthesia for uncooperative patients
- Physical restraint (foam pads, wrapping)
- Swap PE direction to move ghosts away from area of interest
- Increase NSA/NEX (averages) - reduces random motion effects
- Cardiac gating (ECG-triggered or peripheral pulse)
- Respiratory gating (bellows, navigator echo - PACE technique)
- Ultrafast sequences (HASTE, EPI, TrueFISP) - freeze motion in 2-5 seconds
- PROPELLER/BLADE technique - oversample k-space center, detect and reject bad data
- Rest slabs / saturation bands over moving structures
- Antiperistaltic agents (e.g., Buscopan for bowel)
- Respiratory compensation (ROPE, COPE algorithms)
2. Nyquist Ghost (N/2 Ghost)
Appearance: A shifted ghost displaced exactly half the FOV (N/2) in the phase-encode direction, typically seen in Echo Planar Imaging (EPI).
Mechanism:
- EPI acquires all k-space lines in a single TR using alternating gradient polarity (each odd line is read left-to-right, each even line right-to-left).
- Any timing errors, eddy currents, or B₀ inhomogeneity cause a phase offset between even and odd echo lines.
- This even-odd phase inconsistency creates a N/2 frequency modulation in k-space.
- After Fourier transform, this appears as a ghost shifted exactly FOV/2.
Reduction:
- Reference scan (navigator scan without phase encoding) acquired before each EPI acquisition to measure and correct the phase offset
- Improved gradient hardware (faster rise times, better eddy current pre-compensation)
- Ghost correction algorithms in reconstruction software
- Parallel imaging (GRAPPA, SENSE) reduces the number of EPI echoes needed
3. Flow Artifact / Pulsatile Flow Artifact
Appearance: Ghosting in the PE direction from pulsating vessels; may appear as bright or dark signal in vessel lumen.
Mechanism:
- Flowing protons experience different RF pulses and gradient pulses than stationary tissue.
- Inflow enhancement (Time-of-flight effect): fully magnetized fresh spins entering the slice produce brighter signal than partially saturated stationary tissue.
- Pulsatile flow: the velocity of arterial blood changes with the cardiac cycle. Different k-space lines are acquired at different phases of systole/diastole, producing phase inconsistency → ghosts in PE direction.
- Phase dispersion: spins moving through a gradient acquire phase relative to stationary spins. If velocities vary across the vessel, phase spread causes signal loss (especially in GRE).
Reduction:
- Cardiac gating (ECG-trigger data acquisition to same cardiac phase)
- Saturation bands (REST slabs) placed superior/inferior to slice to pre-saturate flowing blood
- Flow compensation (gradient moment nulling / GMN): adds additional gradient lobes that rephase first-order (velocity) motion
- Swap PE direction to redirect ghosts
- Use spin echo rather than GRE sequences (SE refocuses phase errors)
- VENC (velocity encoding) matched appropriately in phase-contrast MRI
CATEGORY 2 - K-SPACE / FOURIER ARTIFACTS
4. Gibbs Ringing / Truncation Artifact
Appearance: Parallel bands of alternating bright and dark signal adjacent to high-contrast interfaces (e.g., brain-CSF border, cord-CSF). Often misidentified as a cord lesion (pseudo-syrinx on sagittal spine).
Mechanism:
- The Fourier transform requires an infinite number of frequencies to represent a sharp discontinuity (step function) perfectly.
- When k-space is truncated (limited matrix size, especially in PE direction), high-spatial-frequency data is absent.
- The incomplete Fourier series produces the Gibbs phenomenon: oscillations (ringing) near the edge.
- The number of rings = half the number of k-space lines.
- More prominent in PE direction because fewer PE steps are used to shorten scan time.
Reduction:
- Increase matrix size (more phase encoding steps) - directly reduces artifact
- Increase FOV in PE direction (oversampling)
- Apply k-space filtering (Hanning window, Fermi filter) - smooths edges but slightly blurs image
- Zero-filling / zero padding in k-space (interpolation, not true resolution improvement)
5. Zipper Artifact
Appearance: A bright line or stripe across the image, oriented in the frequency-encode direction (perpendicular to PE direction), at a specific frequency.
Mechanism:
- External RF interference enters the scanner room (improperly shielded Faraday cage, RF door left open, electronic equipment inside).
- This extraneous RF signal at a frequency within the receiver bandwidth is treated by the system as if it originated from a specific spatial location (determined by the gradient at that moment).
- The system plots it as a line of false signal at that frequency-encode position.
Reduction:
- Ensure RF shielding (Faraday cage) is intact - door must be fully closed during scanning
- Remove electronic devices from scanner room
- Identify and eliminate RF-emitting sources
- Hardware inspection if persistent
6. Herringbone / Crisscross Artifact
Appearance: A series of lines crossing the entire image in a herringbone or corduroy pattern.
Mechanism:
- A single corrupted k-space data point (spike) from a hardware malfunction, electronic transient, or data spike in raw data.
- Because k-space and image space are Fourier pairs, a single delta function (spike) anywhere in k-space transforms to a sinusoidal wave spanning the entire image.
- Multiple spikes or a corrupted k-space line generates a superimposed grid pattern.
Reduction:
- Identify and remove/replace the corrupted k-space line (spike correction algorithms)
- Repeat the sequence
- Hardware maintenance (gradient amplifier, ADC board checks)
- Raw data filtering
7. Data Clipping Artifact
Appearance: Bright areas or signal overflow, image looks "burned out," sometimes with ringing superimposed.
Mechanism:
- The received MRI signal is amplified and digitized by the Analog-to-Digital Converter (ADC).
- If the signal amplitude exceeds the dynamic range of the ADC (receiver gain set too high), the signal is "clipped" - the waveform is truncated at the maximum value.
- This is equivalent to multiplying the true signal by a rectangular window in the time domain, which Fourier transforms to convolution with a sinc function → ringing and edge artifacts.
- Also occurs with incorrectly set prescan gain.
Reduction:
- Proper prescan calibration (auto-prescan)
- Reduce receiver gain (manual correction)
- Do not place high-signal structures (e.g., pads soaked in contrast material) inside the FOV unnecessarily
CATEGORY 3 - FREQUENCY/PHASE ENCODING ARTIFACTS
8. Chemical Shift Artifact (Type 1)
Appearance: A dark band on one side and a bright band on the opposite side of fat-water interfaces (kidney cortex/fat, orbit posterior globe, disc-vertebra interface). Seen in the frequency-encode direction.
Mechanism:
- Water and fat protons precess at slightly different Larmor frequencies due to electron shielding differences.
- At 1.5T: fat resonates ~220 Hz lower than water (3.5 ppm difference × 63.87 MHz = ~224 Hz).
- At 3T: the difference is ~440 Hz.
- The frequency-encoding gradient is used by the system to assign spatial location by frequency.
- Fat signal at a given location is assigned to the wrong pixel (shifted by ΔHz / gradient amplitude in Hz/mm).
- At fat/water interface: where they overlap → bright band; where gap appears → dark band.
Pixel shift formula: Δx = Δf / (BW/pixel)
Where BW = receiver bandwidth per pixel.
Reduction:
- Increase receiver bandwidth (reduces pixel shift, but also reduces SNR)
- Fat suppression (FATSAT, STIR, Dixon) - eliminates fat signal entirely
- Swap to phase-encode direction (shifts artifact but doesn't eliminate)
- Use higher bandwidth sequences
- Avoid narrow bandwidth sequences near fat-water interfaces
9. Chemical Shift Artifact (Type 2) / India Ink / Black Boundary Artifact
Appearance: Black outline around fat-containing structures on GRE sequences. Seen in both FE and PE directions.
Mechanism:
- In GRE sequences, water and fat spins are initially in phase at TE=0.
- Because fat precesses 3.5 ppm slower, they progressively dephase.
- At specific TEs (called "out-of-phase" TEs), water and fat spins in the same voxel point in exactly opposite directions and cancel each other.
- At 1.5T: in-phase TE ~4.6 ms, 9.2 ms; out-of-phase TE ~2.3 ms, 6.9 ms.
- At 3T: in-phase TE ~2.3 ms; out-of-phase TE ~1.15 ms.
- Voxels at fat-water interfaces containing both fat and water protons lose signal → black rim.
Reduction (or deliberate exploitation):
- Choose in-phase TE values (avoid out-of-phase TEs) to minimize artifact
- Out-of-phase imaging is diagnostically used to detect intracellular lipid (adrenal adenoma, fatty liver)
- Dixon technique uses both in-phase and out-of-phase images to separately map fat and water
10. Aliasing / Wrap-Around / Foldover Artifact
Appearance: Anatomy from outside the FOV appears wrapped to the opposite side of the image, superimposed on the structure being imaged.
Mechanism:
- The Nyquist sampling theorem requires the sampling rate to be at least twice the highest frequency present.
- In the PE direction, the sampling rate is determined by the number of PE steps × TR.
- If anatomy extends beyond the FOV, it contains spatial frequencies exceeding the Nyquist limit.
- These aliased frequencies are indistinguishable from lower frequencies and are plotted on the opposite side of the image.
- In FE direction: oversampling is trivial (zero cost), so FE aliasing is automatically suppressed.
- PE direction aliasing is the problem because adding PE steps increases scan time.
Reduction:
- Increase FOV in PE direction (encompasses all anatomy)
- Phase oversampling (No Phase Wrap / anti-aliasing): acquires more PE steps beyond FOV but discards them from display
- Swap PE to a direction where anatomy doesn't extend beyond FOV
- Saturation bands on tissue outside FOV
- Parallel imaging methods
11. Fat-Water Swap Artifact
Appearance: In Dixon fat/water separated images, fat and water assignments are reversed in a region - tissues that should be bright on water image appear bright on fat image and vice versa.
Mechanism:
- Dixon imaging generates fat-only and water-only images by acquiring two echoes (in-phase and out-of-phase) and computing: Water = (IP + OOP)/2; Fat = (IP - OOP)/2.
- Phase errors from B₀ inhomogeneity cause fat and water phases to be misidentified.
- The algorithm applies a region-growing or phase unwrapping algorithm to determine which is fat and which is water; if the algorithm fails in a region (e.g., near air-tissue interfaces), it assigns the labels incorrectly.
Reduction:
- Improved B₀ shimming before scan
- Use 3-point or multi-point Dixon (better phase unwrapping)
- Apply region of interest constraints
- Manual post-processing correction
CATEGORY 4 - SUSCEPTIBILITY / FIELD-RELATED ARTIFACTS
12. Magnetic Susceptibility Artifact
Appearance: Signal void (dropout), geometric distortion, bright/dark halo around metallic objects, dental amalgam, surgical clips, hemorrhage sites. Much worse on GRE sequences than SE.
Mechanism:
- Magnetic susceptibility (χ) describes how much a material becomes magnetized in B₀.
- Ferromagnetic/paramagnetic materials (metal, hemosiderin, air-tissue interfaces) have χ very different from surrounding tissue.
- This creates local B₀ field inhomogeneity (ΔB).
- Local field distortion causes:
a) Frequency shift → spatial misregistration (wrong pixel assignment)
b) Phase dispersion within a voxel → signal loss (T2* decay)
c) GRE sequences are far more sensitive because no 180° refocusing pulse is applied.
d) SE sequences partially compensate via 180° pulse (refocuses static dephasing).
- The "blooming effect" is an exaggeration of the true size of susceptibility sources (hemosiderin, calcification, air).
Reduction:
- Use spin echo instead of gradient echo
- Lower field strength (artifact scales with B₀)
- Shorter TE
- Increase readout bandwidth
- Metal Artifact Reduction Sequences (MARS): view-angle tilting, SEMAC, MARS-SE
- Titanium implants (lower susceptibility than stainless steel or cobalt-chromium)
- STIR rather than CHESS fat suppression near metal
13. Blooming Artifact
Appearance: The area of susceptibility-induced signal loss appears much larger than the actual structure (e.g., a 2mm hemosiderin deposit looks like 8mm signal void on GRE).
Mechanism:
- Extension of susceptibility artifact where the field perturbation extends beyond the physical boundaries of the susceptibility source.
- T2* decay from dephasing spreads signal void beyond the true dimensions.
- Clinically useful: makes microhemorrhages, calcifications, air emboli visible that would otherwise be invisible.
- SWI (Susceptibility Weighted Imaging) deliberately maximizes blooming for hemosiderin detection.
Reduction/Exploitation:
- Reduce by using SE sequences (see susceptibility above)
- Deliberately exploit with GRE/SWI to detect microbleeds in traumatic brain injury, cavernomas, amyloid angiopathy
14. Dielectric Artifact / Dielectric Effect
Appearance: Central brightening or dark bands in the middle of large patients or large structures, particularly at 3T. Seen as regions of signal non-uniformity.
Mechanism:
- The RF wavelength (λ = c / f) at 3T (128 MHz) in tissue is approximately 26 cm.
- When the patient's body dimension approaches or exceeds λ/2, standing wave interference patterns form inside the body.
- This causes constructive interference (bright regions) in the center and destructive interference (dark regions) at the periphery, or vice versa.
- Called "dielectric effect" because tissue permittivity influences wave propagation.
- Also contributes to non-uniform flip angles across the FOV.
- More severe at 3T and 7T than 1.5T.
Reduction:
- Dielectric pads (bags of high-permittivity gel placed on patient) - alter the standing wave pattern
- RF transmit optimization (B1+ shimming with multi-channel transmit coil)
- 3D RF pulses for better B1 uniformity
- Work at 1.5T for large patients
15. Eddy Currents Artifact
Appearance: Geometric distortion, image shearing, ghost artifacts, baseline distortion, particularly in EPI and diffusion sequences.
Mechanism:
- Rapidly switching gradient fields induce currents in conducting parts of the scanner (cryostat, RF shields, gradient coil housing) by Faraday's law of induction.
- These induced currents generate their own secondary magnetic fields that oppose the intended gradient (Lenz's law).
- Eddy fields decay with their own time constants (τ).
- If not corrected, they cause: incorrect slice positions, image warping, phase errors in EPI.
- In diffusion imaging: eddy currents from strong diffusion gradients cause image-to-image geometric distortion that varies with diffusion direction.
Reduction:
- Active gradient shielding (shielded gradient coils): secondary winding cancels eddy field
- Pre-emphasis (pre-distortion) of gradient waveforms: gradient amplifier adds corrective pre-current
- Hardware: improved cryostat design with reduced eddy conductance
- Post-processing: eddy current correction in diffusion imaging (FSL eddy tool, FEE)
- Use of shorter, more efficient gradient pulses (VERSE)
CATEGORY 5 - PARALLEL IMAGING / ACCELERATION ARTIFACTS
16. Parallel Imaging Artifact (Residual Aliasing / g-factor Noise)
Appearance: Residual ghost overlying the image (SENSE artifact), or corduroy pattern noise in specific regions (GRAPPA artifact), particularly in areas with low coil sensitivity variation.
Mechanism:
- Parallel imaging undersamples k-space (acquires every Rth PE line, where R = acceleration factor).
- Undersampling causes R-fold aliasing by Nyquist rules.
- SENSE: uses coil sensitivity maps to "unfold" aliased images. Where coil sensitivity profiles are poorly differentiated (far from coils), unfolding fails → residual aliasing.
- GRAPPA: uses k-space autocalibration signal (ACS) lines to synthesize missing k-space. Errors in fitting produce noise amplification characterized by g-factor.
- Higher R → shorter scan time but worse SNR and more potential artifact.
SNR in parallel imaging: SNR_PI = SNR_full / (g × √R)
Reduction:
- Reduce acceleration factor R
- Use coils with more elements (better coil sensitivity variation)
- Improve ACS calibration (more ACS lines for GRAPPA)
- Regularization in reconstruction (SENSE-REGULARIZED)
- Use combined GRAPPA + SENSE or compressed sensing
17. Moire Fringes
Appearance: Parallel bright and dark fringes across the image, resembling an interference pattern. Usually caused by B₀ inhomogeneity or aliasing interaction.
Mechanism:
- When two periodic signals with slightly different spatial frequencies superimpose, a low-frequency "beat" pattern appears - this is the Moire fringe.
- In MRI, occurs when aliased signal from outside FOV interacts with in-FOV signal, or from wrap-around in conjunction with field inhomogeneity.
- Also seen in sequences where periodic k-space sampling interacts with oscillating physiological signals.
- Can occur in large FOV coronal/axial acquisitions where signal from both sides of the body wraps.
Reduction:
- Phase oversampling to prevent wrap-around
- Increase FOV
- Improved B₀ shimming
- Surface coils with limited coverage
CATEGORY 6 - GEOMETRIC / SPATIAL ARTIFACTS
18. Gradient Non-Linearity Artifact
Appearance: Geometric distortion at the periphery of the image, structures appear "stretched" or "curved" at image edges. Straight lines appear bowed.
Mechanism:
- Ideal gradient fields must be perfectly linear (G = constant × position) for correct spatial encoding.
- Real gradient coils only achieve perfect linearity near the isocenter.
- Toward the edges of the gradient FOV, the gradient field deviates from linearity.
- Non-linear gradients assign incorrect spatial positions to peripheral voxels.
- Objects near the edge appear displaced or distorted (barrel or pincushion distortion).
- Particularly significant in stereotactic planning and radiation therapy target definition.
Reduction:
- 3D gradient warp correction (GradWarp) applied in post-processing using known gradient coil geometry
- Scan at isocenter (center of gradient linear region)
- Gradient coil design improvements (larger linear FOV)
- Dedicated gradient linearity calibration phantoms
19. Stairstep / Staircase Artifact (2D Multislice)
Appearance: On reformatted 2D multiplanar reconstructions (MPR) from multislice 2D acquisitions, curved surfaces appear stepped ("staircase" pattern) when viewed in non-acquired planes.
Mechanism:
- 2D multislice acquisitions have discrete, thick slices with gaps between them.
- When reformatted in an orthogonal plane, the finite slice thickness creates a blocky stepped appearance rather than a smooth curve.
- Unlike isotropic 3D acquisitions, 2D slices have inherently anisotropic voxels (slice thickness >> in-plane resolution).
- Reformat resolution in the z-direction is limited to slice thickness.
Reduction:
- Use thin slices with small or zero gaps
- Use 3D volumetric acquisition (isotropic voxels → smooth reformats in any plane)
- Overlap slices (gap = negative)
- Post-processing interpolation (though adds apparent not true resolution)
20. Partial Volume Artifact
Appearance: Apparent blurring of tissue boundaries; a voxel containing two tissue types appears as intermediate signal, potentially mimicking pathology (e.g., a thin cartilage layer appears intermediate rather than distinctly high signal).
Mechanism:
- MRI assigns a single signal value to each voxel.
- If a voxel contains two tissues (e.g., fat and tumor, or fluid and soft tissue), the signal is the volume-weighted average.
- Thick slices and large in-plane voxels increase the probability of partial volume.
- A large voxel spanning a tissue edge will register the average of both tissues.
- Small lesions smaller than voxel size may be completely obscured.
Reduction:
- Use thinner slices
- Increase in-plane matrix (smaller voxels)
- 3D acquisition with isotropic voxels
- Align slices perpendicular to the interface of interest
- Avoid interpolation as a substitute for true resolution
CATEGORY 7 - SEQUENCE-SPECIFIC ARTIFACTS
21. Magic Angle Artifact
Appearance: Tendons and ligaments (normally very dark on T1 and short-TE sequences due to short T2) appear artificially bright when oriented at approximately 55° to B₀.
Mechanism:
- In ordered structures (collagen fibers in tendon, ligament), the T2 relaxation is highly anisotropic.
- The dipolar coupling between adjacent protons in collagen depends on the orientation of the fiber to B₀.
- The dipolar Hamiltonian contains a (3cos²θ - 1) term. When θ = 54.74° (the "magic angle"), this term = 0.
- At the magic angle, dipolar coupling vanishes, T2 effectively lengthens dramatically (from ~1 ms to ~10-20 ms).
- On sequences with short TE (<20 ms): T1W, PD, GRE - the tendon appears bright.
- On long TE T2W sequences, the extended T2 still decays before signal acquisition → no artifact.
Clinical importance: Can mimic tendon pathology (partial tear, tendinopathy) in shoulder rotator cuff, posterior tibial tendon, Achilles.
Reduction:
- Use long TE sequences (T2W): artifact disappears
- Reposition the body part to change fiber orientation
- Be aware at 55° to B₀ (typically curved tendons passing around pulleys)
22. Cross-Talk / Slice Cross-Excitation Artifact
Appearance: Reduced signal in adjacent slices in 2D multislice acquisitions, appearing as alternating bright-dark banding between slices. Saturation of tissue in adjacent slices.
Mechanism:
- Slice selection RF pulses are not perfectly rectangular in frequency profile - they have imperfect edges and side lobes.
- Adjacent slices share overlapping RF excitation if the slice gap is insufficient.
- A proton in the "tail" of slice 1's profile is partially excited, and before it recovers (T1 recovery), it is excited again by slice 2.
- This partial saturation reduces signal in overlapping regions.
- More severe with closely spaced or contiguous slices.
Reduction:
- Use interleaved slice acquisition (acquire odd slices first, then even - separates simultaneously acquired neighbors)
- Maintain adequate slice gap (typically 10-20% of slice thickness)
- Use 3D volumetric sequences (no inter-slice cross-talk)
- Sinc RF pulses with better slice profiles (or optimized pulses)
23. Projection Artifact / Incomplete Saturation
Appearance: Bright overlapping images of structures outside the intended plane, superimposed on the desired anatomy - appearing like a "projection radiograph" overlay.
Mechanism:
- In 2D imaging, the slice-selective gradient limits excitation to a chosen plane.
- If out-of-plane signal is not properly suppressed (inadequate spoiling, incomplete saturation), residual magnetization from surrounding tissue contributes signal.
- Particularly seen when FOV encompasses high-signal structures (e.g., bowel with oral contrast, fat) that "project" through the slice.
- Can also occur in 3D sequences at slab boundaries.
Reduction:
- Ensure adequate spoiling (gradient spoilers, RF spoiling)
- Saturation bands placed superior/inferior to slab
- Adequate TR to allow T1 recovery discrimination
- Correct slab thickness and position
24. Surface Coil Intensity Variation / Surface Coil Flare
Appearance: Extremely bright signal near the surface coil that fades dramatically with distance, creating non-uniform image illumination. Superficial tissues appear "burned out" while deep structures appear dark.
Mechanism:
- Surface coils are receive-only coils with a spatially varying sensitivity profile.
- SNR and signal intensity decay approximately as 1/r³ from the coil center.
- Near the coil: very high signal (bright).
- Away from coil: rapidly decreasing sensitivity.
- The non-uniform B₁⁻ receive sensitivity is not compensated unless explicitly corrected.
Reduction:
- Intensity normalization (prescan normalization, N3/N4 correction) - post-processing algorithms
- Sensitivity encoding correction (divide image by coil sensitivity map)
- Use volume coils (more uniform) for structures requiring uniformity
- Normalization filters in scanner software
25. Saturation Band Artifact / Dark Band
Appearance: Bands of signal loss where saturation pulses (REST slabs, fat sat) have reduced or eliminated tissue signal unintentionally, appearing as dark lines or zones across the image.
Mechanism:
- Pre-saturation pulses (90° pulses followed by crusher gradients) are applied to suppress signal from specific regions (superior/inferior to imaging slab, fat, vessels).
- If the saturation band overlaps with the imaging volume due to poor planning, anatomy in the overlap zone is saturated and appears dark.
- Similarly, inversion recovery pulses (STIR, FLAIR) null signal at a specific TI; tissues with similar T1 to null-point will be suppressed.
- Dark band artifact also seen in balanced SSFP (bSSFP/TrueFISP) sequences from banding artifacts at points where B₀ causes off-resonance phase of exactly 180°.
Reduction:
- Careful placement of saturation bands (avoid overlap with ROI)
- Review saturation band positions on localizer before prescribing
- For bSSFP banding: improved B₀ shimming, phase-cycling techniques, or use alternative sequences
26. Signal Loss Artifact
Appearance: Regions of absent signal (black areas) where signal should be present.
Mechanism - multiple causes:
- Susceptibility dephasing (see above)
- Poor coil placement - receiver coil too far from ROI
- Through-plane motion during echo time
- Flow void - fast-flowing blood dephases
- T2 decay* from iron deposition, hemosiderin, air
- Inadequate flip angle at that location (B1 non-uniformity)
Reduction: Depends on cause - address the specific source identified above.
CATEGORY 8 - POST-PROCESSING / RECONSTRUCTION ARTIFACTS
27. Gibbs/Ringing in Image Stitching / Composing Artifact
Appearance: Visible seam lines or intensity jumps where multiple image acquisitions are combined to cover a large FOV (e.g., whole-spine imaging, whole-body DWI).
Mechanism:
- Large anatomy (full spine, legs) requires multiple overlapping acquisitions with different coil positions/stations.
- Differences in B₁ uniformity, coil sensitivity, patient position shift between stations, or normalization errors create intensity mismatches at the stitch boundaries.
- Automated blending algorithms attempt to smooth transitions but residual seam lines may remain.
Reduction:
- Adequate overlap between stations
- Consistent normalization across stations
- Table-feed (continuous table movement) acquisitions (whole-body MRI) reduce station boundaries
- Use system-integrated AutoAlign or CARE protocols
28. Spatially Dependent Phase Artifact
Appearance: Apparent signal changes that vary spatially and do not correspond to tissue properties - phase images show unexpected phase distributions.
Mechanism:
- B₀ inhomogeneity across the FOV causes spatially varying resonance frequency offsets.
- During phase-sensitive acquisitions (phase contrast MRA, susceptibility mapping, Dixon), the background phase from B₀ variation adds to the desired phase signal.
- This "background phase" is spatially dependent and can mimic or obscure the signal of interest.
Reduction:
- High-order B₀ shimming before acquisition
- Background phase correction (polynomial fit and subtraction)
- Reference phase acquisition for subtraction
- PREBASELINE correction algorithms
CATEGORY 9 - CONTRAST AGENT ARTIFACTS
29. Gadolinium Pseudo-Layering Artifact
Appearance: On post-contrast sequences, a false "fluid-fluid level" or layering appearance within a cyst or joint space not corresponding to true anatomical layering.
Mechanism:
- Gadolinium chelates diffuse into fluid-containing spaces and may concentrate at the dependent portion due to their higher density.
- Gd shortens T1 of adjacent tissue/fluid; concentrated Gd creates T1 shortening at a specific depth.
- On T1W images this appears as a layer of bright signal resembling a fluid level.
- Can mimic intra-articular loose bodies, hemorrhagic layering, or complex cyst.
Reduction:
- Clinical correlation - recognize the phenomenon
- Repeat imaging after repositioning patient (true fluid levels shift; pseudo-layering also shifts but differently)
- Delay post-contrast timing
30. Radiation-Induced Artifact (MR-Linac / Hybrid MRI)
Appearance: Artifacts in hybrid MR-radiotherapy (MR-Linac) systems - ghosting, geometric distortion, noise patterns caused by the interplay between radiation beam delivery and simultaneous MRI acquisition.
Mechanism:
- In MR-guided radiotherapy (MRgRT), MRI is acquired simultaneously with radiation delivery.
- The radiation beam interacts with the main magnetic field - secondary electron trajectories are deflected by the Lorentz force (F = qv × B).
- The radiation ionizes gas and generates RF emission that contaminates MRI signal.
- Gradient eddy currents from rapidly switching MR gradients affect dosimetry.
Reduction:
- Temporal interleaving of MRI acquisition and beam delivery
- RF shielding optimization within the MR-Linac bore
- Hardware filtering; specialized reconstruction algorithms
- Sequence optimization for MR-Linac environment
CATEGORY 10 - ADDITIONAL ARTIFACTS
31. Signal-to-Noise Ratio (SNR) Artifacts
Appearance: "Grainy" or "noisy" image with poor contrast between structures.
Physics of SNR:
SNR ∝ (Voxel volume) × √(Number of averages × acquisition time) / (BW)^0.5
Factors reducing SNR:
- Small voxels (thin slice, high matrix, small FOV)
- High receiver bandwidth
- Low field strength
- Poor coil-patient coupling
- High body temperature (Johnson noise)
Improvement:
- Larger voxels (compromise resolution)
- More averages (longer scan time)
- Narrow bandwidth (more chemical shift artifact)
- Better coil design
- Higher field strength
32. Starry Sky Artifact
Appearance: Multiple small bright foci scattered across the image, resembling stars against a dark background. Typically seen on DWI or T2* sequences.
Mechanism:
- Phase encoding errors from multiple small motion events, gradient imperfections, or timing inconsistencies.
- In EPI-DWI: each bright "star" is caused by a localized eddy current or motion event affecting a single or few k-space lines.
- Corrupt k-space lines → point spread function artifacts.
Reduction:
- EPI correction
- Navigator echoes to detect and reject corrupted shots
- Multi-shot DWI with motion correction
33. Lipid Suppression Failure Artifact
Appearance: Bright fat signal persisting in areas where fat suppression was applied, creating heterogeneous suppression (bright fat in some regions, suppressed elsewhere).
Mechanism:
- CHESS (Chemical Shift Selective) fat saturation uses a frequency-selective 90° pulse tuned to the fat resonance (3.5 ppm from water).
- B₀ inhomogeneity shifts the local resonance frequency away from the CHESS pulse center frequency.
- In regions where B₀ is inhomogeneous (near air-tissue interfaces, metallic implants, curved surfaces), fat is no longer at the expected frequency → CHESS pulse fails to saturate fat.
- Water protons may instead be saturated, reducing water signal.
Reduction:
- Shimming before fat-sat sequences
- STIR (Short Tau Inversion Recovery): T1-based fat suppression - not frequency dependent → more robust near metal and at field edges
- Spectral STIR or water-only excitation (SPAIR, IDEAL)
- Dixon-based fat separation (most robust, frequency-independent)
34. The "Annefact" / Antenna Effect
Appearance: Signal hyperintensity at tissue-air interfaces, particularly in patients with long conductive structures (e.g., pacemaker leads, wires) that act as antennas.
Mechanism:
- Conductive wires resonant with the RF transmit frequency accumulate RF energy (antenna effect).
- The wire tip acts as a high-impedance source → local tissue heating AND local B₁ enhancement → abnormal signal.
- Length of resonance: ~λ/2 at Larmor frequency (at 1.5T: ~26 cm for half-wavelength antenna).
Safety relevance: Also a patient safety concern as tip heating can cause burns.
Reduction:
- Avoid long conductive leads inside the scanner without MR-conditional clearance
- Use MR-conditional pacemaker protocols
- Reduce RF power (SAR)
- Position leads to minimize resonant coupling
35. Zebra Stripe Artifact
Appearance: Alternating black and white stripes across the image in a zebra-like pattern, typically in bSSFP (True FISP, FIESTA) sequences.
Mechanism:
- bSSFP sequences maintain a steady-state transverse magnetization between TR cycles.
- The steady-state signal is critically dependent on the resonance offset frequency.
- At off-resonance frequencies of ±1/(2TR), the steady-state signal drops to zero (banding nulls).
- B₀ inhomogeneity causes spatial variation in resonance offset → alternating bright (on-resonance) and dark (off-resonance/banding) bands.
- Band spacing = 1/TR in frequency; physically separated by distance proportional to B₀ gradient.
Reduction:
- Very short TR (pushes banding nulls farther apart in Hz - wider "pass band")
- High-order B₀ shimming
- Phase cycling methods (rotate RF phase to shift banding locations, combine images)
- Use non-bSSFP alternative if banding unacceptable
36. BLADE/PROPELLER Residual Artifact
Appearance: Subtle blurring or residual motion ghosts in BLADE/PROPELLER acquisitions despite motion correction - sometimes over-corrected with slightly reduced detail.
Mechanism:
- BLADE (Siemens)/PROPELLER (GE) acquires k-space as rotating blades/propellers, oversampling the k-space center.
- The iterative registration algorithm detects and corrects in-plane rotation and translation.
- If motion is through-plane, or if the blade itself contains motion, the correction is incomplete.
- Over-correction can over-smooth the final image.
Reduction:
- Motion during data acquisition minimized (patient coaching)
- Increased number of blades (more overdetermination for correction)
- BLADE-specific reconstruction parameters (weighting of outlier blades)
SUMMARY TABLE - QUICK REVISION
| Artifact | Direction | Cause | Key Fix |
|---|
| Motion ghosting | PE direction | Patient/physiological motion | Gating, fast sequences, swap PE |
| Nyquist ghost | PE (FOV/2) | EPI even/odd line phase error | Reference scan correction |
| Chemical shift 1 | FE direction | Fat-water frequency difference | Increase BW, fat suppression |
| Chemical shift 2 | All directions | In/out-of-phase GRE | Choose correct TE |
| Aliasing/wrap | PE direction | FOV too small | Phase oversampling, increase FOV |
| Gibbs/truncation | PE direction | Small matrix, k-space truncation | Increase PE steps |
| Susceptibility | All (worse GRE) | B₀ distortion near metal/air | Use SE, short TE, MARS |
| Zipper | FE direction (line) | RF interference | Fix Faraday cage |
| Herringbone | Whole image | k-space spike | Spike correction, hardware fix |
| Magic angle | N/A | 55° fiber orientation | Use long TE (T2W) |
| Gibbs/parallel | Whole image | Inadequate coil sensitivity | Reduce R, better coil |
| Cross-talk | Slice direction | Adjacent slice excitation overlap | Interleaved acquisition, gaps |
| Dielectric | Central/peripheral | Standing RF waves at 3T | Dielectric pads, B1 shimming |
| Eddy current | PE + FE | Induced currents from switching gradients | Shielded gradients, pre-emphasis |
| Fat-water swap | Whole image | Phase error in Dixon | Better shimming, multi-point Dixon |
| Partial volume | Slice direction | Large voxels spanning 2 tissues | Thin slices, 3D acquisition |
| Gradient non-linearity | Peripheral FOV | Non-linear gradients at edge | GradWarp correction, scan at isocenter |
| Surface coil flare | Near coil | Non-uniform coil sensitivity | Normalization correction |
| Stairstep | z-direction | 2D thick slice MPR | 3D isotropic acquisition |
| bSSFP banding / Zebra | All | B₀ off-resonance in SSFP | Short TR, shimming, phase cycling |
| Lipid suppression failure | Near metal/edges | B₀ inhomogeneity disrupting CHESS | STIR, Dixon fat suppression |
| Blooming | Around susceptibility source | T2* dephasing extension | SE sequences, short TE |
| Data clipping | Whole image | ADC overflow | Reduce receiver gain |
HIGH-YIELD EXAM POINTS
- Chemical shift type 1 - frequency-encode direction, increases with lower bandwidth, increases with field strength (3T worse than 1.5T)
- Gibbs ringing - pseudo-syrinx in spine - increase PE steps or use k-space filter
- Magic angle - always use T2W to confirm/deny tendon pathology
- Aliasing - PE direction only (FE oversampled automatically)
- Susceptibility - GRE >> SE; scales with B₀; STIR better than CHESS near metal
- Nyquist ghost - EPI specific; FOV/2 displacement; corrected by reference scan
- Dielectric artifact - 3T-specific; central brightening; dielectric pads help
- Zipper - always FE direction; always RF contamination; fix the shielding
- bSSFP banding - short TR is the primary solution (moves nulls to wider frequency separation)
- Parallel imaging noise - g-factor amplification; reduce R or use more coil elements
- Cummings Otolaryngology Head and Neck Surgery, pp. 149-150
- Bradley and Daroff's Neurology in Clinical Practice (GRE susceptibility, SWI sections)
- MRI Questions - Motion Artifacts
- Medical Professionals - MRI Artifact Guide