when is a pediatric circuit used, like in what age patients

Reading File
Finding Sources
Finding Sources
Searching PubMed

"pediatric" AND "breathing circuit"

Reading File
Reading File
Reading File
Searching the Web

pediatric breathing circuit vs adult circuit weight cutoff kg age anesthesia ventilator neonatal pediatric circuit

A pediatric circuit is chosen by weight and size more than by age. In practice it means infants and small children, up to about 10 to 30 kg depending on the type of circuit and the local protocol.
Rough guide
PatientTypical circuit
Neonates and infants, under about 10 kg (roughly under 1 year)Low-resistance, low-dead-space circuits: Mapleson D, Bain, or Jackson-Rees (Mapleson F), or a pediatric circle
Toddlers and young children, about 10 to 20 kgPediatric circle or Jackson-Rees
Children over about 20 to 30 kg (roughly 6 to 8 years and up)Adult circle system
The weight and age ranges here are approximate and vary by source. They are rules of thumb, not fixed cutoffs.
What the textbooks say
  • Morgan and Mikhail's Clinical Anesthesiology, 7e says that for patients under 10 kg, some anesthesiologists prefer the Mapleson D or Bain system. These have low resistance and are light. It also says the circle system can be used safely at any age if ventilation is controlled or assisted. The concern is a sick neonate breathing spontaneously, who may struggle against the resistance of the valves, tubing and absorber.
  • Barash, Cullen, and Stoelting's Clinical Anesthesia, 9e describes the Jackson-Rees (Mapleson F) circuit as "commonly used in pediatrics." It has low resistance and minimal dead space. It needs a fresh gas flow of about 2.5 times minute ventilation to prevent rebreathing. Compared with a pediatric circle, it may cause slightly less difficulty in breathing and less gastric insufflation.
  • Morgan and Mikhail's Clinical Anesthesiology, 7e adds that pediatric circle systems often have a septum in the Y-piece and low-compliance tubing. Both reduce dead space and the tidal volume lost to circuit compliance. They are also lighter.
What the other sources add
  • The WFSA paediatric equipment guide says it is customary to use adult breathing systems in children over 20 to 30 kg.
  • For ICU and transport ventilators, the cutoffs are based on weight and tidal volume. For example, Fisher & Paykel recommends changing from an infant to an adult circuit at about 12 to 15 kg or a tidal volume of about 120 mL. Hamilton Medical's pediatric circuit sets cover about 3 to 15 kg. Always check the manufacturer's labeling for your equipment.
  • The Pediatric Anesthesia Digital Handbook matches the reservoir bag to the child. A 1 L bag is used for about 6 months to 6 years (under 25 to 30 kg), and a 2 L bag for older children (about 30 to 50 kg).
Why use a smaller circuit
  • It lowers dead space, so there is less CO2 rebreathing in small tidal volumes.
  • Its compliance is lower, so less delivered tidal volume is lost to the tubing.
  • Its resistance is lower, which matters for a spontaneously breathing infant.
  • It is lighter, so there is less drag on a small endotracheal tube.
The PubMed check for recent meta-analyses or systematic reviews on this topic found nothing that changes the above.

What should be the fresh gas flow