First: what an ECG actually measures
An ECG does not measure the heart’s pumping force, and it does not normally measure the amount of electrical current flowing through the patient.
It measures tiny voltage differences at the skin produced by the summed electrical activity of heart cells. These voltage differences are usually measured in millivolts (mV).
The ECG machine compares the electrical potential sensed at two locations, amplifies the difference, filters unwanted noise, and draws the result as the P wave, QRS complex, and T wave.
1. Volt and millivolt (mV): the main ECG unit
Meaning
A volt (V) is a unit of electrical potential difference, also called voltage.
A millivolt (mV) is one-thousandth of a volt:
[
1\text{ mV} = 0.001\text{ V}
]
The electrical signals reaching the skin from the heart are very small, commonly fractions of a millivolt to a few millivolts. Therefore, ECGs use mV rather than V.
What voltage means in an ECG
Think of voltage as the electrical “push” or difference in potential between two points.
For example, Lead I compares the potential at the left arm with that at the right arm:
[
\text{Lead I} = V_{\text{left arm}} - V_{\text{right arm}}
]
If the left-arm sensing site is more positive than the right-arm site, the ECG trace deflects upward in Lead I. If it is more negative, the trace goes downward.
Standard ECG calibration
A conventional ECG is recorded at:
[
\boxed{1\text{ mV} = 10\text{ mm}}
]
This is called standard gain or standardization.
So, on the ECG paper:
| ECG height | Actual voltage |
|---|
| 1 mm, one small vertical square | 0.1 mV |
| 5 mm | 0.5 mV |
| 10 mm, two large vertical squares | 1 mV |
| 20 mm | 2 mV |
Example 1: QRS amplitude
Suppose the R wave in Lead II is 15 mm high and the S wave is 3 mm deep.
- R wave = (+15) mm = (+1.5) mV
- S wave = (-3) mm = (-0.3) mV
- Net QRS amplitude = (1.5 - 0.3 = +1.2) mV
At standard calibration, the ECG’s 1 mV calibration pulse should produce a vertical rectangle 10 mm tall. Standard clinical recordings commonly use 25 mm/s paper speed and 10 mm/mV gain, as described in
this ECG protocol.
Why mV matters clinically
Voltage measurements help assess findings such as:
- Low-voltage QRS complexes, which can occur in conditions such as large pericardial effusion, obesity, COPD/emphysema, or hypothyroidism. Interpretation always needs clinical context.
- High QRS voltage, one possible ECG feature of ventricular hypertrophy.
- ST-segment elevation or depression, measured in millimetres or mV from the isoelectric baseline.
For instance, at standard gain:
[
1\text{ mm ST elevation} = 0.1\text{ mV}
]
2. Ampere (A): electrical current
Meaning
An ampere (A) is the unit of electric current, meaning the rate at which electric charge moves:
[
I = \frac{Q}{t}
]
Where:
- (I) = current in amperes
- (Q) = charge in coulombs
- (t) = time in seconds
Common smaller units are:
[
1\text{ mA} = 0.001\text{ A}
]
[
1\text{ µA} = 0.000001\text{ A}
]
What current means in the heart
Heart cells generate electrical activity because ions move across cell membranes:
- Sodium ((\text{Na}^+))
- Calcium ((\text{Ca}^{2+}))
- Potassium ((\text{K}^+))
- Chloride ((\text{Cl}^-))
These ionic currents create changing voltage differences across cardiac cell membranes. When many cardiac cells activate together, their summed electrical field can be detected on the body surface as an ECG.
Does an ECG machine send current into the patient?
A standard diagnostic ECG is primarily a passive recording device. It senses voltage differences from the body. It is not intended to drive a clinically meaningful current through the heart.
This differs from:
- Pacemaker: deliberately delivers electrical impulses to stimulate the heart.
- Defibrillator: delivers a large, controlled electrical shock to terminate certain dangerous arrhythmias.
- Electrocautery/diathermy: intentionally passes high-frequency current through tissue.
- ECG: records the heart’s naturally occurring surface voltage signal.
Some ECG systems may use extremely small internal currents for technical functions such as checking electrode contact or reducing electrical interference, but this is not the cardiac signal being interpreted.
Example 2: why current is not printed on the ECG
A QRS complex may be recorded as 1 mV, but the ECG paper does not tell you “the heart current is X amperes.”
Why not?
Because the measured voltage depends on many factors:
- direction of the heart’s electrical vector
- position of the heart in the chest
- tissue conductivity
- skin and electrode contact
- body size
- distance from heart to electrodes
- the particular lead being viewed
Thus, ECG interpretation focuses on voltage, time, wave shape, and lead distribution, rather than estimating current in amperes.
3. Ohm (Ω): resistance and impedance
Meaning
An ohm (Ω) is the unit of electrical resistance.
Resistance describes how much a material opposes current flow. The basic relationship is Ohm’s law:
[
\boxed{V = I \times R}
]
Where:
- (V) = voltage in volts
- (I) = current in amperes
- (R) = resistance in ohms
Rearrangements:
[
I = \frac{V}{R}
]
[
R = \frac{V}{I}
]
Simple numerical example
If 1 V is applied across a resistor of 1000 Ω:
[
I = \frac{V}{R} = \frac{1}{1000} = 0.001\text{ A}
]
[
= 1\text{ mA}
]
This is a basic electrical example. It does not mean that an ECG applies 1 V through the patient.
4. Resistance versus impedance in ECG
For direct current, “resistance” is the usual term. ECG signals change continuously with time, so the more accurate term is often impedance.
Impedance includes resistance plus the frequency-dependent opposition created by components such as skin and electrode gel.
In ECG practice, people may say “electrode resistance,” “lead resistance,” or “electrode impedance.” In a practical sense, they are referring to how well the electrode-skin connection permits accurate sensing of the tiny ECG voltage.
The electrode-skin interface
The heart’s electrical signal must pass through:
[
\text{heart} \rightarrow \text{body tissues} \rightarrow \text{skin} \rightarrow \text{gel} \rightarrow \text{electrode} \rightarrow \text{lead wire} \rightarrow \text{ECG machine}
]
The skin-electrode interface is often the biggest practical source of poor signal quality.
High or unequal electrode impedance can cause:
- baseline wander
- movement artifact
- 50/60-Hz mains interference
- intermittent or noisy trace
- distortion of small waves such as P waves
- unreliable ST-segment measurement
5. Example: why skin preparation improves an ECG
Dry, oily, hairy, sweaty, or poorly cleaned skin may create a poor electrode connection and higher impedance.
Practical steps
- Explain the procedure and ensure the patient is relaxed.
- Expose the required areas.
- Clean oil, sweat, and lotion from the skin.
- Dry the area.
- Clip excessive hair if it prevents adhesive contact. Do not shave more than needed.
- Apply fresh electrodes firmly to intact skin.
- Ask the patient to lie still, avoid talking, and relax the arms and legs.
Better contact lowers impedance and reduces artifact.
Example 3: artifact from poor contact
A patient is shivering or moving their arm. The electrode moves slightly over the skin. The contact impedance varies repeatedly, and the ECG baseline becomes irregular or wavy. This may resemble arrhythmia, but it is artifact.
The response is not to diagnose an arrhythmia immediately. First check:
- patient movement or shivering
- loose electrode
- dried-out electrode gel
- poor skin contact
- cable movement
- nearby electrical interference
6. How voltage, current, and resistance relate to ECG
| Quantity | Unit | Meaning in ECG |
|---|
| Voltage | volt, usually mV | The primary ECG measurement: electrical-potential difference detected between electrode sites |
| Current | ampere, mA, µA | Ionic current in heart cells generates the electrical field; the ECG does not normally display a current measurement |
| Resistance/impedance | ohm, kilo-ohm (kΩ) | Opposition at the skin-electrode interface; high or imbalanced values degrade tracing quality |
A useful analogy
- Voltage: water pressure difference
- Current: amount of water flowing per second
- Resistance: narrowing or blockage of the pipe
- ECG: mainly measures the “pressure difference” at different points on the body surface, not the amount of water flowing through the pipe.
7. Why an ECG machine needs a high input impedance
The ECG voltage is very small. The machine must sense it without drawing enough current to alter or weaken the signal.
Therefore, ECG amplifiers are designed with very high input impedance. In simple terms, the machine “listens” to the voltage signal rather than loading the circuit.
If the machine’s input impedance were low, it could draw more current from the sensing circuit and potentially distort the recorded voltage.
8. The 1 mV calibration mark
Before or during an ECG, the machine may show a rectangular calibration signal.
At normal gain:
[
\boxed{1\text{ mV calibration signal} = 10\text{ mm vertical height}}
]
This lets the clinician know whether waveform amplitudes can be interpreted correctly.
Example 4: altered gain
Sometimes the voltage is so high that the QRS complexes run off the paper. The operator may use half-standard gain:
[
1\text{ mV} = 5\text{ mm}
]
Then a 10-mm-high R wave does not equal 1 mV. It equals 2 mV.
Conversely, if signals are very small, the ECG may be amplified more, for example:
[
1\text{ mV} = 20\text{ mm}
]
Always inspect the calibration label before measuring amplitudes.
Key points to remember
- Millivolt (mV) is the central ECG amplitude unit.
- At normal calibration, 1 mV equals 10 mm vertically on ECG paper.
- Ampere measures current. Cardiac ionic currents cause electrical activity, but a routine ECG mainly records resulting surface voltage differences.
- Ohm (Ω) measures resistance. In ECG, skin-electrode impedance strongly affects signal quality.
- Good skin contact, correct electrode placement, patient stillness, and proper calibration are necessary for a reliable ECG.
- An ECG is a voltage-recording device, unlike a defibrillator or pacemaker, which intentionally deliver electrical energy.
The standard bipolar limb-lead recording concept and the voltage relationship between leads are described in Guyton and Hall Textbook of Medical Physiology, p. 146.