Note on Baheti and Dorsch & Dorsch: These two texts are not available in the library database, but their content has been retrieved via the Baheti textbook PDF (Understanding Anesthetic Equipment & Procedures, Dwarkadas K. Baheti et al.) and cross-referenced with the PMC article. Dorsch & Dorsch (Understanding Anesthesia Equipment, 5e) is referenced in both Miller's and the PMC article but is not in the digital library; key content from its protocols is captured below via those references.
Pre-Anaesthesia Equipment Check and Machine Maintenance Protocols
(Sources: Miller's Anesthesia 10e; Barash's Clinical Anesthesia 9e; Morgan & Mikhail's Clinical Anesthesiology 7e; Baheti - Understanding Anesthetic Equipment & Procedures; PMC3821271 [Indian Journal of Anaesthesia - Comprehensive PAC Protocol]; ASA Recommendations for Pre-Anaesthesia Checkout, 2008)
I. Background and Rationale
Improperly checking anaesthesia equipment prior to use can lead to patient injury, severe morbidity, and mortality. The
ASA Closed Claims database shows equipment-related claims have fallen to approximately 1% of all claims - partly attributed to mandated PAC procedures. However, available evidence consistently shows that anaesthesia providers frequently do not perform a complete pre-anaesthesia check (PAC) and may miss faults even when explicitly looking for them on a sabotaged machine. In 35% of equipment-related adverse events, a proper PAC would have prevented the outcome.
Key facts (Miller's Anesthesia, 10e):
- A complete PAC must be performed each day before first use
- An abbreviated version must be performed before each subsequent case
- All contemporary anesthesia workstations have automated checkout procedures - but none can fully replace a manual checkout; automated checks differ between models and cannot be assumed to cover all critical items
II. Governing Guidelines and Standards
| Organisation | Standard / Guideline |
|---|
| ASA (USA) | Recommendations for Pre-Anesthesia Checkout Procedures (2008) - 15-item framework |
| FDA (USA) | Anesthesia Apparatus Checkout Recommendations (now superseded by ASA 2008) |
| AAGBI (UK) | Checking Anaesthetic Equipment (2012) - applicable to all workstations |
| ISO | ISO 80601-2-13:2011 - Safety and essential performance of anaesthetic workstations |
| ASTM International | F1850-00 - Standard specifications for anaesthesia machines |
Baheti (Understanding Anesthetic Equipment): Reproduces both the ASA 2008 15-item PAC and the AAGBI checklist as Appendices 1 and 2, noting the AAGBI checklist is applicable to all workstations and should be used alongside manufacturer-specific checks.
III. ASA 2008 Recommendations: The 15-Item PAC Framework
(Barash 9e, Appendix B; Miller's 10e, Table 20.6; Baheti, Appendix 1 & 2)
Items to be Completed DAILY (Before First Case)
| # | Item | Responsible Party | Rationale |
|---|
| 1 | Verify auxiliary O₂ cylinder and self-inflating manual ventilation device (Ambu bag) are available and functioning | Provider AND technician | Failure to ventilate is a major cause of anaesthesia morbidity/mortality; must be present at every anaesthetising location |
| 3 | Turn on anaesthesia delivery system; confirm AC power is available | Provider OR technician | Machine may run on battery backup silently - first sign of failure can be catastrophic system shutdown when battery is exhausted |
| 5 | Verify pressure is adequate on the spare O₂ cylinder mounted on the machine | Provider AND technician | Backup O₂ must be available if pipeline fails; valve should be closed after checking to prevent inadvertent emptying |
| 6 | Verify piped gas pressures ≥50 psig | Provider AND technician | Minimum supply pressure is required for proper machine function |
| 8 | Verify no leaks in the gas supply lines between flowmeters and the common gas outlet | Provider OR technician | The low-pressure section (flowmeters → vaporizers → common gas outlet) is most vulnerable; each vaporizer must be turned ON individually during leak test |
| 9 | Test scavenging system function | Provider OR technician | Protects OR personnel from waste anaesthetic gases |
| 10 | Calibrate (or verify calibration of) the O₂ monitor; check low-O₂ alarm | Provider OR technician | Final defence against hypoxic gas delivery |
Items to be Completed BEFORE EACH CASE
| # | Item | Responsible Party |
|---|
| 2 | Verify patient suction adequate to clear the airway | Provider AND technician |
| 4 | Verify availability of required monitors; check alarms (SpO₂, NIBP, capnography, ECG) | Provider OR technician |
| 7 | Verify vaporizers adequately filled; filler ports tightly closed | Provider ONLY |
| 11 | Verify CO₂ absorbent is not exhausted | Provider OR technician |
| 12 | Breathing system pressure and leak testing | Provider AND technician |
| 13 | Verify gas flows properly through the breathing circuit (inspiration and exhalation) | Provider AND technician |
| 14 | Document completion of checkout procedures (on patient chart) | Provider AND technician |
| 15 | Confirm ventilator settings; ANAESTHESIA TIME-OUT | Provider ONLY |
IV. Detailed Protocol by System: Pneumatic, Electrical, Electronic Components
(Synthesised from Miller's 10e; Morgan & Mikhail 7e [Table 4-3, FDA Checkout]; Baheti Appendix 2 [Ravishankar Protocol]; PMC3821271)
A. High-Pressure System
- Oxygen cylinder: Open O₂ cylinder; verify it is at least half full (≥1000 psig) (Morgan) or gauge pressure adequate (Miller's). Then close the cylinder valve - if left open and pipeline fails, the cylinder depletes silently.
- All gas cylinders: Check N₂O, air and other cylinders only if those gases are required for the case.
- Central pipeline supplies: Confirm all hoses are connected and pipeline gauges read ~50 psig.
- Colour coding and DISS connections: Visually inspect; perform a "tug test" (AAGBI/Baheti) to confirm correct insertion of each hose into its Schrader socket. (Note: Repeated disconnection of hoses to test the O₂ failure alarm may cause premature failure of the Schrader socket - the AAGBI therefore recommends checking the O₂ failure alarm weekly by disconnecting rather than daily.)
B. Low-Pressure System (Most Vulnerable to Leaks)
- Initial status check: Close all flow control valves; turn vaporizers off; check vaporizer fill levels and tighten filler caps.
- Negative-pressure leak test (for machines with a common gas outlet check valve):
- Attach suction bulb to the fresh gas outlet
- Squeeze bulb repeatedly until fully collapsed
- Verify bulb stays collapsed for ≥10 seconds (detects leaks as small as 30 mL/min)
- Repeat with each vaporizer turned on individually (to check vaporizer mount seals)
- Remove bulb; reconnect fresh gas hose
- Positive-pressure leak test (for machines without a check valve):
- Can use back-pressure through the breathing circuit
- Vaporizers must still be tested individually
- Vaporizer filler ports: Partially open filler ports are a common leak source. Automated checkout typically does NOT test for vaporizer leaks unless the vaporizer is opened during the test.
C. Flowmeters
- Turn on master switch and all electrical equipment.
- Adjust the flow of all gases through their full range; check for smooth operation of floats and undamaged flow tubes.
- Hypoxic guard test: Attempt to create a hypoxic O₂:N₂O mixture; verify the correct changes in flow or alarm activation (proportioning system check).
- Confirm minimum O₂ flow: With O₂ flow control valve off, verify O₂ flowmeter registers ~50-200 mL/min (minimum mandatory flow).
D. Scavenging System
- Ensure proper connections between the scavenging system and both the APL (pop-off) valve and the ventilator relief valve.
- Adjust waste gas vacuum if adjustable.
- Under-pressure test: With minimum O₂ flow, fully open APL valve, occlude Y-piece, allow scavenger reservoir bag to collapse completely - absorber pressure gauge should read ~zero.
- Over-pressure test: With O₂ flush activated, allow scavenger bag to distend fully - absorber pressure gauge should read <10 cmH₂O.
E. Breathing System / Circle System
- O₂ analyser calibration:
- Expose sensor to room air - should read 21%; if not, calibrate
- Verify low-O₂ alarm is enabled and functioning
- Reinstall sensor and flush with O₂ - should read >90%
- Circuit integrity: Check that breathing circuit is complete, undamaged, and unobstructed; verify unidirectional valves are present and moving correctly.
- CO₂ absorbent: Verify it is not exhausted (colour change indicates exhaustion).
- Leak test of breathing system:
- Set all gas flows to zero; close APL (pop-off) valve; occlude Y-piece
- Pressurize circuit to ~30 cmH₂O with O₂ flush
- Ensure pressure remains fixed for ≥10 seconds
- Open APL valve and verify pressure decreases
- Accessory equipment: Install humidifier, PEEP valve, and other circuit accessories planned for the case.
F. Ventilator
- Place a test lung (or second breathing bag) on the Y-piece.
- Set appropriate ventilator parameters for the next patient.
- Switch to automatic mode; turn ventilator on.
- Fill the bellows and cycle the ventilator; observe bellows function and unidirectional valves.
- Confirm minimum volume alarm is enabled and functioning.
- Check for adequate ventilation of test lung and confirm absence of leaks.
- Return to manual/bag mode before the patient is connected.
G. Monitors
- Check SpO₂, NIBP, ECG, ETCO₂, temperature probe, and nerve stimulator as appropriate.
- Self-test: The Baheti/Ravishankar protocol recommends using monitors on oneself (e.g., SpO₂ on own finger >96%; exhale into capnograph port to generate a CO₂ waveform) before applying to the patient.
- Set alarm limits appropriate for the patient's profile.
- Confirm gas sampling lines are properly attached, patent, and free from kinks.
H. Electrical/Electronic Components (Baheti / Ravishankar Protocol)
- Confirm machine is connected to mains (AC) and switch is on.
- Ensure battery has ≥30 min backup and is charging during machine use.
- For desflurane vaporizers: confirm electrical power and power-up self-test of the vaporizer.
V. The MS MAIDS Mnemonic (Miller's Anesthesia, 10e - Anesthesia Time-Out)
Before each case, an "Anaesthesia Time-Out" can be performed using this mnemonic (Box 20.3, Miller's 10e):
| Letter | Component |
|---|
| M - Machine | PAC complete; vaporizers filled, closed, dial at "0"; all gas flow knobs at zero; ventilator settings appropriate; APL valve open; machine in manual/spontaneous mode |
| S - Suction | Patient suction adequate to clear airway |
| M - Monitors | All required standard monitors present and ready |
| A - Airway | Primary airway equipment AND appropriate backup ready |
| I - Intravenous | IV lines, fluids, and associated equipment ready |
| D - Drugs | All medications available and properly labelled |
| S - Special | Any special or unique items (additional monitors, equipment) for the case are available |
VI. AAGBI Checklist Principles (Baheti, Appendix 1)
- Applicable to all anaesthetic workstations
- Should take only a few minutes to perform
- Is not intended to replace the manufacturer's pre-anaesthetic checks - use alongside them
- For machines with automated self-testing cycles on switch-on: functions tested by the machine need not be manually retested, but the operator must confirm what functions are covered
- The balance: "not so superficial that its value is doubtful, nor so detailed that it is impractical to use"
- O₂ failure alarm: The AAGBI recommends checking this on a weekly basis by disconnecting the O₂ hose (not daily), to avoid premature failure of the Schrader socket; a "tug test" is the daily standard
VII. Maintenance Protocols
Preventive Maintenance
- The PAC is not a replacement for required preventive maintenance (Barash, General Considerations)
- Anaesthesia machines should undergo scheduled professional servicing by manufacturer-certified biomedical technicians
- Clear documentation of regular servicing, component replacement, and satisfactory post-service functioning is mandatory (PMC3821271)
- Vaporizers should be serviced and recalibrated at manufacturer-recommended intervals (typically every 1-2 years)
Between Cases
- Test all new equipment intended for the next patient (suction tubing, breathing circuitry)
- Verify sufficient CO₂ absorbent and volatile agent
- If O₂ cylinder was used during a case: confirm it is at least half full or replace with a full cylinder
- During long procedures: periodically check for exhaustion of volatile agent and CO₂ absorbent
Handover During Long Cases (PMC3821271)
- If anaesthesiologist changes during a long procedure, a formal handover of machine check status must be given, including all component functioning
Machine Relocation
- Whenever an anaesthesia machine is moved to a new location, a complete beginning-of-day checkout must be performed (Barash, Basic Principles)
VIII. Limitations of Automated Checkout (Miller's 10e)
Automated checkout procedures:
- Differ between manufacturers and models
- Cannot be assumed to check all required items - reading the user manual does not always clarify what is being tested
- May not test for leaks at the vaporizer specifically, so loose filler caps or vaporizer mount leaks can be missed
- Many providers do not fully understand what automated checks cover, or make false assumptions about their machine's self-test
- Automated checks should clearly distinguish what is automatically tested from what requires manual checkout
- Ideally, date, time, and outcome of the most recent check should be recorded and accessible to the user
IX. Documentation
- Whoever conducts the PAC must document successful performance; the anaesthesia provider must include this on the patient chart (Barash, Basic Principles)
- Documentation serves as a quality control tool and medicolegal record
- Training in the PAC should also be formally documented, including competency items and successful completion
X. Key Exam Points
- The ASA 2008 Recommendations have 15 items - 15 items daily, 8 items before each subsequent case (items 2, 4, 7, 11-15)
- Item 7 (vaporizer check) and Item 15 (Anaesthesia Time-Out) are the only items exclusively the provider's responsibility - technicians cannot substitute
- The negative-pressure test (suction bulb method) is used on machines with a common gas outlet check valve; the positive-pressure test cannot be used on these machines
- Vaporizer leak test: each vaporizer must be turned on individually - automated checkout typically misses vaporizer leaks
- O₂ failure alarm: test weekly by disconnecting O₂ hose; use a "tug test" daily (AAGBI/Baheti)
- The PAC should be in an ergonomically rational sequence and as short as possible without omitting critical items
- Dorsch & Dorsch (Understanding Anesthesia Equipment, 5e) is the classic reference for in-depth equipment checkout procedures and forms the basis of many institution-specific protocols, though the ASA 2008 guidelines are the current operative standard
Primary references: Miller's Anesthesia, 10e (Chapter 20, pp. 2400-2423); Barash's Clinical Anesthesia, 9e (Appendix B, pp. 2071-2086); Morgan & Mikhail's Clinical Anesthesiology, 7e (Table 4-3, pp. 146-148); Baheti DK et al., Understanding Anesthetic Equipment & Procedures, Appendices 1 & 2; Ravishankar M - PMC3821271