Now I have comprehensive information from multiple textbooks. Here is a thorough explanation:
For example, sounds may be heard at 180 mmHg (true systole), disappear between 160 and 140 mmHg, then reappear at 140 mmHg, and finally disappear again at 90 mmHg (diastole). The gap in this case is 160–140 mmHg.
-
Arterial stiffness — Most commonly seen in hypertensive patients and the elderly, stiff, non-compliant arteries alter the turbulent flow dynamics that generate Korotkoff sounds. As the cuff is deflated, turbulent flow sufficient to produce audible sounds may not be consistently generated at intermediate pressure ranges.
-
Impaired venous return / venous congestion — When venous return to the arterial side is reduced (e.g., from venous congestion in the arm, or the cuff itself compressing venous outflow), the reduced arterial filling can cause a transient drop in the intensity of Korotkoff sounds to below the audible threshold. Once the cuff pressure falls further, enough blood flows through to again generate turbulence and sound.
-
Exaggerated vasomotor tone or peripheral vasoconstriction — Vasoconstriction dampens the turbulence needed to produce sounds.
-
Underestimation of systolic BP: If the clinician starts inflating the cuff at a low pressure (without first palpating the radial pulse to determine true systolic), they may begin hearing sounds only after the gap has passed — and wrongly record this as the systolic pressure. The gap can account for up to a 25 mmHg discrepancy. (The Washington Manual of Medical Therapeutics)
-
Incorrect diastolic BP: The reappearance of sounds after the gap may be mistaken for a new phase, and the gap may be misinterpreted as the diastolic pressure.
-
Palpate the radial pulse first — Inflate the cuff until the radial pulse disappears; this gives a rough systolic estimate. Inflate 20–30 mmHg above this level before auscultating. This prevents starting auscultation within or below the gap.
-
Deflate slowly — 2–3 mmHg per second.
-
Auscultate all the way to zero — Do not stop at the first point of silence; continue listening until the manometer approaches zero to avoid mistaking the gap for diastole. (Roberts and Hedges' Clinical Procedures in Emergency Medicine)
-
Use digital/oscillometric devices — Automated monitors measure oscillations and are not subject to the auscultatory gap.
Sources: Fuster and Hurst's The Heart, 15e | The Washington Manual of Medical Therapeutics | Morgan & Mikhail's Clinical Anesthesiology, 7e | Roberts and Hedges' Clinical Procedures in Emergency Medicine | Swanson's Family Medicine Review