I now have comprehensive content from Braunwald's chapter 69. This is the full CIED chapter. I have everything I need to construct a thorough, residency/DM-level note. Let me now compose the comprehensive notes:Here are comprehensive, residency/DM-level notes on Pacemakers and Intracardiac Devices - built from base physiology to advanced clinical management.
Pacemakers and Intracardiac Devices
Cardiology Residency / DM Level Notes
1. CLASSIFICATION OF CARDIAC IMPLANTABLE ELECTRICAL DEVICES (CIEDs)
CIEDs deliver therapeutic electrical stimuli and encompass:
| Device | Stimulus type | Primary purpose |
|---|
| Permanent pacemaker (PPM) | Low voltage (1-5 V) pacing pulses | Bradycardia |
| Implantable Cardioverter-Defibrillator (ICD) | High voltage (500-1400 V) shocks + pacing | VF/VT termination + backup pacing |
| CRT-P (biventricular pacemaker) | LV + RV pacing pulses | Cardiac resynchronization in HF |
| CRT-D (biventricular ICD) | Same as above + defibrillation | Resynchronization + SCD prevention |
| Leadless pacemaker | Intracardiac capsule, no lead | Single-chamber pacing, no pocket |
| Subcutaneous ICD (S-ICD) | Subcutaneous coil + electrode | VF/VT termination without intravascular lead |
| Loop recorder (ILR) | Passive sensing only | Long-term arrhythmia monitoring |
Source: Braunwald's Heart Disease, Chapter 69 (Mina K. Chung & James P. Daubert)
2. ANATOMY OF A PACING SYSTEM
Pulse Generator (PG)
- Contains battery (lithium iodide; ~5-12 years lifespan) + circuitry
- "End of life" (EOL) signaled by pacing rate or magnet response change (typically rate-responsive or VOO/DOO switch)
- Programmable via telemetry (radiofrequency wand or Bluetooth)
Lead System
- 1-3 leads depending on device type
- Lead body: conducting wires + insulating material (polyurethane or silicone)
- Fixation mechanisms:
- Active fixation: extendable/retractable screw (helical tip) - can be placed anywhere, used in atrium, septum, His bundle, LV epicardium
- Passive fixation: tines/fins (rely on trabecular entrapment) - most common in RV apex
- Lead polarity:
- Unipolar: distal tip (cathode) to generator can (anode) - larger pacing spike, more susceptible to EMI
- Bipolar: distal tip (cathode) to proximal ring electrode (anode) - smaller spike, less EMI susceptibility
Electrode-Tissue Interface
- Cathodal stimulation used for pacing (distal electrode = cathode)
- Strength-duration curve: relates pulse voltage/current to pulse duration for capture
- Rheobase: minimum voltage at infinitely long pulse duration (~1.5-2.0 msec)
- Chronaxie: pulse duration at which threshold = 2x rheobase - this is the most energy-efficient pacing point
- Energy formula: J = V²t(1/R), so optimizing near chronaxie balances safety and battery longevity
- Pacing threshold increases transiently after implantation (acute injury from active fixation), peaks at ~1-2 weeks, then falls ("maturation")
- Output typically programmed at 2x threshold voltage to maintain adequate safety margin
3. NASPE/BPEG PACEMAKER CODE (NBD CODE)
| Position | I | II | III | IV | V |
|---|
| Meaning | Chamber Paced | Chamber Sensed | Response to Sensing | Rate Modulation | Multisite Pacing |
| Options | O, A, V, D | O, A, V, D | O, T, I, D | O, R | O, A, V, D |
- O = None; A = Atrium; V = Ventricle; D = Dual (A+V)
- T = Triggered; I = Inhibited; D = Dual (T+I)
- R = Rate responsive
Common modes explained:
| Mode | Meaning | When used |
|---|
| VVI | Ventricular pacing, ventricular sensing, inhibited | AF with bradycardia, single-chamber backup |
| VVIR | Above + rate adaptive | AF with chronotropic incompetence |
| DDD | Both paced/sensed, inhibited + triggered | Normal sinus rhythm, AV block |
| DDDR | DDD + rate adaptive | SSS/AV block with activity |
| AAI | Atrial pacing, atrial sensing, inhibited | SSS with intact AV conduction |
| VOO/DOO | Asynchronous - no sensing | Magnet application; EMI-prone setting |
| DDI | Dual sensing/pacing, inhibited only (no tracking) | AF prevention when atrial tracking would be unsafe |
DDD Timing Cycles (Key)
- AV delay (AVD): interval from atrial event to ventricular pacing (analog of PR interval)
- Lower Rate Interval (LRI): max interval between paced beats = 60,000/LRL (msec)
- Upper Rate Limit (URL): maximum tracking rate to prevent 1:1 atrial tracking at fast sinus rates
- PVARP (Post-Ventricular Atrial Refractory Period): atrial channel blanked after V event - prevents sensing retrograde P waves as pacemaker-mediated tachycardia (PMT)
- TARP = AVD + PVARP: Total Atrial Refractory Period - sets Wenckebach upper rate behavior
- Rate-responsive AV delay: shortens AVD at faster rates to mimic physiology
- Safety pacing: if crosstalk (atrial pulse sensed on ventricular channel), device paces ventricle ~100 ms after atrial pacing to prevent inhibition
4. SENSING AND CAPTURE PHYSIOLOGY
Capture
- Successful depolarization of myocardium by pacing stimulus
- Loss of capture on ECG: pacing spike not followed by P or QRS
- Common causes: lead dislodgment, exit block (elevated threshold due to fibrosis, electrolyte abnormalities, drugs like flecainide/amiodarone), lead fracture, output programmed too low, battery depletion
Sensing
- Device detects intrinsic cardiac activity via EGM (intracardiac electrogram)
- Oversensing: device "sees" signals it should not (T waves, far-field R waves, myopotentials, EMI) - results in inappropriate inhibition of pacing (dangerous in pacing-dependent patients) or inappropriate ICD shocks
- Undersensing: device fails to detect intrinsic events - results in asynchronous pacing (competitive rhythm risk), loss of rate-responsive behavior
- Sensitivity setting is a voltage threshold: lower number = more sensitive (detects smaller signals)
5. RATE-RESPONSIVE PACING
- Addresses chronotropic incompetence (heart rate fails to rise appropriately with exertion)
- Sensors used:
- Accelerometer (piezoelectric): detects body movement/vibration - most common
- Minute ventilation: detects thoracic impedance changes with breathing
- QT interval sensor: QT shortens with sympathetic stimulation
- Dual-sensor devices combine above for more physiologic response
- DDDR or VVIR modes; sensor gain/slope programmable
6. HEMODYNAMIC ASPECTS OF PACING
- AV synchrony: atrial kick contributes ~20-30% of cardiac output; loss (ventricular pacing without atrial synchrony) reduces CO, raises atrial pressure -> "pacemaker syndrome" (retrograde VA conduction, neck vein pulsations, dizziness, dyspnea)
- Pacing site matters:
- RV apical pacing: causes dyssynchronous contraction (LBBB pattern) -> adverse remodeling over time, can cause pacing-induced cardiomyopathy
- RV outflow tract (RVOT) / septal pacing: more physiologic
- His bundle pacing (HBP): activates the native conduction system -> narrow QRS, most physiologic; technically challenging, higher thresholds
- Left bundle branch area pacing (LBBAP): emerging technique, more reliable capture than HBP, captures LBB or LBB fascicles deep in septum
7. INDICATIONS FOR PERMANENT PACEMAKERS
Class I (Symptomatic Bradycardia / High-Degree Block)
Sinus Node Dysfunction (SND / SSS):
- Symptomatic sinus bradycardia (including drug-induced when no alternative exists)
- Symptomatic chronotropic incompetence
- Symptomatic sinus pauses
- Tachycardia-bradycardia syndrome with symptomatic bradycardia
AV Block:
- Third-degree (complete) AV block - regardless of symptoms if rate <40 bpm or escape rhythm unreliable
- Second-degree Mobitz II AV block - block below His, high risk of progression to complete block
- Advanced second-degree block (2:1 or higher) with symptoms
- Symptomatic second-degree Mobitz I (Wenckebach) - less common, usually AV nodal
- Complete AV block after acute MI persisting >5-7 days
Bundle Branch Block / Bifascicular block:
- Alternating LBBB and RBBB
- RBBB + LAFB or LPFB with syncope when EP study shows HV ≥100 ms or infra-His block
Post-ablation:
- Permanent AV block after AV node ablation
Hypersensitive Carotid Sinus Syndrome:
- Symptomatic cardioinhibitory carotid sinus hypersensitivity (asystole >3 sec with symptoms)
Neurocardiogenic Syncope:
- Recurrent reflex syncope with documented pauses (DDI or DDD with hysteresis)
Guidelines: 2018 ACC/AHA/HRS Guideline on Bradycardia and Cardiac Conduction Delay (Kusumoto FM et al., Circulation 2019)
8. PACEMAKER TROUBLESHOOTING
Failure to Pace (No Output)
- Lead fracture, connector pin problem, oversensing (inhibition), battery depletion, "safety switch" to ERI mode
Loss of Capture (Output but No Response)
- Lead dislodgment, elevated threshold (drug effect: flecainide, amiodarone; metabolic: hyperkalemia, hypothyroidism; fibrosis), output too low, insulation break
Oversensing
- Causes: T-wave oversensing, far-field R-wave sensing (atrial channel sensing ventricular events), myopotentials (pectoral muscle in unipolar), EMI
- Consequence: inappropriate pacing inhibition (if pacer dependent = syncope); inappropriate ICD shock
- Fix: decrease sensitivity (raise threshold), switch to bipolar, reprogram refractory periods
Undersensing
- Causes: inadequate EGM amplitude (scar, dislodgment), inappropriate sensitivity setting, intrinsic signal timing within blanking period
- Consequence: competitive pacing, loss of inhibition, potential R-on-T phenomenon
Pacemaker-Mediated Tachycardia (PMT)
- Mechanism: retrograde VA conduction after ventricular pacing -> retrograde P wave sensed outside PVARP -> triggers another V pace -> loop
- Rate: upper tracking rate
- Termination: magnet (asynchronous pacing), lengthen PVARP, enable PMT termination algorithm (most modern devices)
Pacemaker Syndrome
- Retrograde VA conduction with VVI pacing -> no AV synchrony -> low CO, cannon A waves, dyspnea, dizziness
- Treatment: upgrade to dual-chamber system
Runaway Pacemaker
- Rare; device malfunction causing very rapid pacing (>200 bpm); battery end-of-life or component failure
- Emergency: magnet application, device re-programming or emergent explant
9. IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS (ICDs)
Device Types
- Transvenous ICD (TV-ICD): lead in RV (sensing/pacing + shocking coil), ± SVC coil; generator in left pectoral pocket
- Single-chamber (VR): RV only
- Dual-chamber (DR): RA + RV leads - better discrimination of SVT vs VT
- CRT-D: RA + RV + LV (CS) leads
- Subcutaneous ICD (S-ICD): no intravascular lead; generator at left axilla; subcutaneous coil + sensing electrode along sternum
- Advantages: avoids transvenous lead complications; for patients with no pacing/CRT need
- Disadvantages: no antibradycardia pacing, no ATP, larger device, inappropriate shocks from T-wave oversensing
- Extravascular ICD (EV-ICD, e.g., EMPOWER): lead tunneled substernally (no intravascular), generator left pectoral; can provide ATP
- Wearable cardioverter-defibrillator (WCD): external vest; temporary protection while deciding on ICD or bridge during waiting period (e.g., newly diagnosed LV dysfunction)
Shock Waveform
- Biphasic waveform: 500-1400 V; reverses polarity mid-shock - reduces defibrillation energy requirement by ~30% vs monophasic
- Defibrillation threshold (DFT): tested at implant (or not, as per contemporary practice); safety margin typically ≥10 J below maximum device output
ICD Therapies
- Antitachycardia Pacing (ATP): sequence of rapid pacing pulses (burst or ramp) to terminate VT without a shock - effective in up to 90% of VTs <250 bpm; painless; first-line therapy for VT
- Low-energy cardioversion: synchronized shock at lower energy; for organized VT
- High-energy shock: defibrillation; for VF or failed cardioversion
- Antibradycardia pacing: backup pacing (post-shock asystole, drug-induced bradycardia)
Detection Zones
- VF zone (typically >200-220 bpm): shock therapy
- VT zone (typically 150-200 bpm): ATP first, then shock if ATP fails
- VT monitor zone (optional): detection + recording only, no therapy
- Detection enhancements (SVT-VT discrimination):
- Onset: sudden onset favors VT; gradual onset favors SVT
- Stability: irregular RR intervals favor AF
- Morphology: EGM morphology compared to template
- AV relationship (dual-chamber): VA dissociation confirms VT
10. INDICATIONS FOR ICD IMPLANTATION
Secondary Prevention (post VF/VT arrest - strongest evidence)
- Survivors of cardiac arrest from VF or hemodynamically unstable VT not due to reversible causes
- Sustained VT with structural heart disease
- Meta-analysis: ICD vs amiodarone -> 50% reduction in arrhythmic death, 25% reduction in all-cause mortality
Primary Prevention
Ischemic cardiomyopathy:
- MADIT-I / MADIT-II / SCD-HeFT trials established benefit
- SCD-HeFT: EF ≤35%, NYHA II-III, optimal medical therapy -> amiodarone vs ICD vs placebo; ICD reduced all-cause mortality 23% vs placebo
- MADIT-II: post-MI, EF ≤30% -> ICD reduced mortality 31%
- Guideline: ICD for LVEF ≤35%, NYHA II-III, ischemic cardiomyopathy, on optimal medical therapy ≥3 months, expected survival >1 year with good functional status
Non-ischemic cardiomyopathy:
- DEFINITE, SCD-HeFT: benefit also shown in NICM with EF ≤35%, NYHA II-III
- DANISH trial (2016): ICD vs control in NICM - no significant reduction in all-cause mortality (HR 0.87, p=0.28), though significant reduction in SCD; contemporary HF therapy may have diluted benefit
- Still indicated by guidelines for NICM, EF ≤35%, NYHA II-III after ≥3 months optimal therapy
Specific cardiomyopathies:
- HCM: ICD for aborted cardiac arrest, spontaneous sustained VT, ≥2 major SCD risk factors (FHx SCD, NSVT, syncope, LV wall thickness ≥30 mm, abnormal BP response to exercise)
- Arrhythmogenic right ventricular cardiomyopathy (ARVC): ICD for aborted arrest, sustained VT; consider for high-risk features
- Brugada syndrome: ICD for aborted VF/VT; quinidine as adjunct; high-risk features (spontaneous type 1 pattern + syncope)
- Long QT syndrome: ICD if aborted arrest or VT on beta-blockers; beta-blockers first line
- Catecholaminergic polymorphic VT (CPVT): nadolol/flecainide first; ICD if VT/VF on therapy
- Sarcoidosis with cardiac involvement: ICD often indicated with EF <35% or significant conduction disease
- Chagas disease: ICD for VT/VF, high-risk features
Guideline threshold for primary prevention:
- EF ≤35% after 3 months optimal medical therapy (regardless of etiology in most guidelines)
- Note: Waiting period is critical - LVEF must be reassessed after 3-6 months of GDMT; early ICD implantation post-MI (within 40 days) has NOT shown benefit (DINAMIT trial)
11. CARDIAC RESYNCHRONIZATION THERAPY (CRT)
Mechanism
- Patients with LBBB (or wide QRS) have dyssynchronous LV contraction (late lateral wall activation) -> reduced LV efficiency
- CRT paces RV + LV simultaneously (biventricular pacing) to restore synchrony -> improved stroke volume, reduced MR, reverse remodeling over time
Indications (Class I)
- LVEF ≤35% + LBBB morphology + QRS ≥150 ms + NYHA II-IV + on GDMT ≥3 months
- NYHA II on GDMT: CRT-D (ICD + CRT) - MADIT-CRT, RAFT trials
- NYHA III-IV: CRT-P or CRT-D - CARE-HF, COMPANION trials
Key Trials
| Trial | Population | Result |
|---|
| CARE-HF | NYHA III-IV, EF≤35%, QRS≥120ms | CRT reduced mortality 36%; CRT-P non-inferior to CRT-D for mortality |
| COMPANION | NYHA III-IV, EF≤35%, QRS≥120ms | CRT-P reduced HF hospitalization + death; CRT-D best |
| MADIT-CRT | NYHA I-II, EF≤30%, QRS≥130ms | CRT-D reduced HF events 41% vs ICD alone; LBBB benefited most |
| RAFT | NYHA II-III, EF≤30%, QRS≥120ms | CRT-D reduced death + HF hospitalization vs ICD |
| ECHO-CRT | NYHA III-IV, EF≤35%, narrow QRS (<130ms) | CRT harmful - increased mortality |
Key teaching point: CRT is harmful in narrow QRS (<130 ms). QRS morphology matters: LBBB gets greatest benefit; RBBB/non-specific IVCD benefit is uncertain.
LV Lead Placement
- LV lead placed via coronary sinus into a lateral or posterolateral branch
- Problem in ~5-10%: cannot reach ideal vein; alternatives include surgical epicardial lead, LV endocardial pacing
- Adequate biventricular pacing: must achieve >98% BiV pacing for optimal benefit
AV Node Ablation + CRT
- When AF + rapid ventricular rate cannot be controlled medically and CRT device is implanted, AV node ablation ensures 100% biventricular pacing
- Rate control alone + CRT: underperforms ablation + CRT in patients needing full resynchronization
12. LEADLESS AND NOVEL DEVICES
Leadless Pacemaker (Micra, Aveir)
- Capsule implanted directly into RV via femoral venous approach; no subcutaneous pocket, no transvenous lead
- Single-chamber: Micra VR (VVIR) - for AF + bradycardia
- Dual-chamber: Micra AV (VDDR) - uses mechanical sensing of atrial contraction to trigger ventricular pacing; Aveir DR - two separate communicating capsules
- Complications: cardiac perforation/tamponade (~1.5%), device embolization, vascular access issues; retrieval is possible but challenging
- Advantages: no pocket (no pocket hematoma/infection), no lead (no lead failure)
Subcutaneous ICD (S-ICD) - Boston Scientific EMBLEM
- No intravascular component; entirely subcutaneous
- Sensing via 3 sensing vectors from subcutaneous electrode
- Pre-implant screening ECG mandatory (SENSE screening) to ensure adequate T:R ratio to avoid T-wave oversensing
- Cannot deliver ATP or chronic pacing - not suitable for pacing-dependent patients, VT requiring ATP, or patients needing CRT
- Particularly preferred in: young patients (preserve vasculature), prior TV-ICD lead infections, no pacing need
Extravascular ICD (EV-ICD)
- Lead tunneled under sternum; generator left pectorally
- Combines benefits of S-ICD (no intravascular) with ability to deliver ATP
- Emerging technology (EFFORTLESS EV-ICD registry)
Conduction System Pacing (CSP)
- His bundle pacing (HBP): electrode placed at His bundle; corrects LBBB by capturing native conduction; narrow QRS
- Challenges: higher threshold, risk of lead dislodgment, sensing issues
- Left bundle branch area pacing (LBBAP) / Left bundle branch pacing (LBBP): electrode screwed deep into interventricular septum to capture LBB directly
- More stable, lower thresholds than HBP, corrects LBBB
- Increasingly used as alternative to biventricular CRT (CSP-CRT or His-optimized CRT)
- LBBP-RESYNC trial: ongoing evaluation vs BVP-CRT
13. COMPLICATIONS OF CIEDs
Procedural
| Complication | Notes |
|---|
| Pneumothorax | ~1-2%; subclavian access > axillary > cephalic approach |
| Hemothorax | Vascular injury; may require drainage |
| Cardiac perforation/tamponade | ~0.5-1%; right ventricular perforation; presents with pleuritic pain, effusion |
| Lead dislodgment | Early (within 24 hrs): most common with atrial lead; Late: fibrosis makes re-positioning difficult |
| Arterial puncture | Inadvertent subclavian artery; direct pressure or surgical repair |
| Air embolism | Venous air entry via dilator; patient Trendelenburg, aspiration |
| Diaphragmatic pacing | LV (CS) lead proximity to phrenic nerve; re-position or reduce output |
Long-term
| Complication | Management |
|---|
| CIED infection | Most serious; involves pocket or leads; TEE to rule out lead vegetation; explant entire system; IV antibiotics 6 weeks; reimplant on contralateral side after 72h negative cultures (if pacing-dependent, temporary pacing bridge) |
| Pocket hematoma | DAPT/anticoagulation related; drain only if expanding/infected; avoid early drainage (increases infection risk) |
| Lead failure | Insulation breach (impedance fall) or conductor fracture (impedance rise); oversensing / undersensing pattern; manage with lead abandonment + new lead |
| Subclavian crush syndrome | Lead fracture at clavicle-rib junction; switch to axillary vein approach |
| Twiddler's syndrome | Patient manipulates generator -> lead rotation -> dislodgment, retraction |
| Venous thrombosis | Subclavian/axillary/SVC; anticoagulation if symptomatic; SVC syndrome rare |
| Tricuspid regurgitation | RV lead interference with TV leaflet; mechanism: impingement, adhesion, perforation |
14. ELECTROMAGNETIC INTERFERENCE (EMI)
| Source | Risk | Management |
|---|
| MRI | Most CIEDs are now MRI-conditional (specific programming required) | Programme to asynchronous mode (VOO/DOO); turn off tachyarrhythmia detection; 1.5T preferred; cardiologist + MRI team coordination |
| Electrosurgery (cautery) | Oversensing -> inhibition or inappropriate shock | Program to asynchronous/suspend detection; use bipolar cautery when possible; donut/ring magnet over ICD inhibits detection during case |
| External defibrillation | Can damage CIED; defibrillator pads placed as far from device as possible (anterior-posterior preferred) | Check device post-shock |
| Lithotripsy | Inhibition/shocks; trigger off R-wave | Suspend ICD detection; lithotripter focus away from device |
| Radiotherapy | Cumulative radiation damage to CIED circuitry | Reposition device if in field; frequent device checks during treatment |
| Mobile phones, airport security | Very low risk with current devices | Maintain 15-cm separation from phone; walk through security without dwelling |
Magnet behavior:
- Over a pacemaker: applies asynchronous (VOO/DOO) pacing at magnet rate (~85 bpm typically; manufacturer-specific) - prevents oversensing inhibition
- Over an ICD: suspends tachyarrhythmia detection (no shock); does NOT change pacing mode
15. ICD TROUBLESHOOTING
Appropriate Shocks
- Delivery of shock for true VT/VF - desired
- If recurrent: optimize anti-arrhythmic drugs (amiodarone, sotalol, mexiletine), catheter ablation for scar-related VT, reassess reversible causes
Inappropriate Shocks (IS)
- Definition: ICD shock delivered for non-VT/VF rhythm
- Most common cause: AF with rapid ventricular rate (most frequent overall), T-wave oversensing, SVT, lead fracture/noise
- T-wave oversensing: more common with S-ICD; manage by adjusting sensing vectors, programming sensitivity
- Lead noise (fracture): large-amplitude, non-physiologic signals; seen as very short RR intervals on EGM; reprogramming or lead replacement
- Consequence of IS: significant psychological distress, possible proarrhythmia (shock on vulnerable period), increased mortality in some analyses
- Management of IS: identify and eliminate cause; antiarrhythmics for AF; shock reduction algorithms; S-ICD screening
VT Storm
-
3 separate episodes of VT requiring ICD intervention within 24 hours
- Management:
- Correct reversible causes (ischemia, electrolytes, drugs)
- Amiodarone IV (150 mg bolus, then infusion)
- Beta-blockers IV (metoprolol, propranolol)
- Deep sedation (reduces sympathetic tone)
- Catheter ablation (urgent if storm persists)
- Temporary overdrive pacing if bradycardia-dependent VT
- Mechanical circulatory support (IABP, Impella) for cardiogenic shock
Failure to Shock / Detect VF
- VF within detection programming (confirm detection turned ON)
- Oversensing causing double-counting (VF detection criteria reached falsely); undersensing of VF
- Elevated DFT (drug effect: amiodarone increases DFT; hypothyroidism)
- Lead fracture, connector issue
16. REMOTE MONITORING
- All modern CIEDs transmit data wirelessly to a base station (home monitor) -> manufacturer server -> clinician portal
- Allows detection of: arrhythmias, lead problems (impedance change), battery status, fluid status (intrathoracic impedance as proxy for pulmonary congestion)
- IN-TIME, CONNECT, TRUST trials: remote monitoring reduces mortality and hospitalization
- Standard of care: remote monitoring + in-office device checks every 6-12 months
17. SPECIAL SITUATIONS
CIED in Pregnancy
- Pacing is safe; lead implantation during pregnancy if urgently needed (use echocardiographic guidance to reduce radiation)
- ICDs can deliver shocks safely to fetus without harm at standard energies
Perioperative CIED Management
- If pacing-dependent: reprogram to asynchronous mode pre-op; have external pacing/defibrillation available
- ICD detection: suspend via magnet or reprogramming; ensure magnet available throughout case
- Postoperatively: restore original settings; check device interrogation
End of Life / CIED Deactivation
- Ethically permissible to deactivate ICD at patient/family request (avoiding unwanted shocks at end of life)
- Pacemaker deactivation more complex (patient may be pacemaker-dependent)
- Requires advance care planning; palliative care team involvement
CIED Extraction (Transvenous Lead Removal - TLE)
- Indications: infection (Class I), lead malfunction affecting clinical management, lead recalled for safety
- Techniques: simple traction, locking stylets, laser sheaths (Spectranetics), mechanical dilator sheaths, femoral workstation
- Risks: vascular/cardiac tear (~1% major complications at high-volume centers), hemothorax, tamponade; requires cardiac surgical backup
18. KEY EXAM/CLINICAL PEARLS
| Point | Detail |
|---|
| Wenckebach (Mobitz I) site | Usually AV node; QRS narrow; usually benign; Atropine helps |
| Mobitz II site | Below His bundle (His-Purkinje); wide QRS; pacemaker mandatory |
| Complete heart block with wide escape | Infranodal; unreliable; pacemaker urgently |
| 2018 guidelines landmark | Bradycardia: Kusumoto et al., Circulation 2019 |
| Waiting 40 days post-MI for ICD | DINAMIT showed no benefit from early ICD (day 6-40 post-MI) |
| S-ICD prerequisite | SENSE screening mandatory pre-implant |
| CRT harm | Narrow QRS <130 ms (ECHO-CRT); always check QRS width |
| CRT best responders | LBBB + QRS ≥150 ms + ischemic CM + women |
| Most common IS cause | AF with rapid rate; then T-wave oversensing |
| Magnet on pacemaker | Asynchronous pacing at magnet rate (device-dependent, ~85 bpm) |
| Magnet on ICD | Suspends detection only; does NOT pace asynchronously (unless also bradycardic) |
| PMT termination | Lengthen PVARP; PMT algorithm delivers AV refractory extension |
| Pacemaker syndrome | VVI mode + retrograde VA conduction; upgrade to DDD |
| Tricuspid regurgitation from lead | Lead removal + TV repair if severe |
| Most feared CIED complication | CIED infection; needs complete system removal |
| DFT increase | Amiodarone, hypothyroidism, hyper/hypokalemia, antiarrhythmic drugs |
Sources
- Braunwald's Heart Disease, Chapter 69 - Mina K. Chung & James P. Daubert (CIEDs)
- Braunwald's Heart Disease, Chapter 68 (Bradyarrhythmias and Conduction Disorders)
- Fuster and Hurst's The Heart, 15th Edition, Chapter 37 (Ventricular Arrhythmias/SCD) and Chapter 36 (AF)
- Referenced guidelines: ACC/AHA/HRS 2018 Bradycardia Guideline; AHA/ACC/HRS 2017 VT/SCD Guideline; ESC 2021 Pacing GuidelinesThese are complete, residency/DM-level notes built from base to top. Here's a quick orientation to the structure:
The 18 sections progress logically:
- Device classification (what each CIED is)
- Hardware anatomy (PG, leads, electrodes)
- NBD pacing code (understand any programmed mode)
- Sensing & capture physics (strength-duration curve, chronaxie, threshold maturation)
- Rate-responsive pacing (sensors, chronotropic incompetence)
- Hemodynamics of pacing (AV synchrony, pacing site effects, His/LBBAP)
- Pacemaker indications (SND, AV block, bifascicular, hypersensitive carotid - Class I/II)
- Pacemaker troubleshooting (failure to pace, loss of capture, over/undersensing, PMT, runaway pacemaker)
- ICD types (TV-ICD, S-ICD, EV-ICD, WCD) + shock physics
- ICD therapy zones (ATP, cardioversion, defibrillation, detection zones)
- ICD indications (secondary prevention meta-analysis; primary prevention trials: MADIT, SCD-HeFT, DANISH, with key LVEF/QRS thresholds)
- CRT (mechanism, indications, key trials with data: CARE-HF, COMPANION, MADIT-CRT, RAFT, ECHO-CRT)
- Leadless, S-ICD, CSP/HBP/LBBP (emerging tech)
- Complications (procedural + long-term including infection management)
- EMI (MRI, cautery, magnet behavior)
- ICD troubleshooting (inappropriate shocks, VT storm algorithm)
- Remote monitoring, special situations (pregnancy, perioperative, end-of-life, TLE)
- High-yield exam pearls (table format)
All content sourced from Braunwald's Heart Disease (9th ed.) and Fuster & Hurst's The Heart (15th ed.).