Anesthesia machine check list

Reading File
Finding Sources
Searching the Web

anesthesia machine pre-use checklist ASA 2008 steps

Finding Sources
Reading Web Page

https://aneskey.com/machine-checkout-and-quality-assurance

Reading File
Reading File
Reading File
Reading File
Reading File
I now have comprehensive data from two leading anesthesia textbooks (Miller's Anesthesia 10e and Morgan & Mikhail's Clinical Anesthesiology 7e) plus the Canadian Anesthesiologists' Society 2022 guidelines. Here is the full compiled checklist:

Anesthesia Machine Pre-Use Checklist

Based on the 2008 ASA Recommendations for Pre-Anesthesia Checkout (PAC), as detailed in Miller's Anesthesia 10e and Morgan & Mikhail's Clinical Anesthesiology 7e.
Note: Because modern anesthesia workstations differ widely, no single procedural checklist applies to all machines. Each department should develop a machine-specific version based on these principles and the manufacturer's manual. Items marked * must be completed before every case, not just daily.

FREQUENCY KEY

  • D = Daily (or when machine is moved / vaporizer changed)
  • E = Before every case

SECTION 1 - Emergency Backup (D)

Item 1: Verify Backup Ventilation & Oxygen
  • Confirm a self-inflating manual resuscitation bag (e.g., Ambu bag) is present and functional at every anesthetizing location
  • A non-self-inflating Mapleson circuit is NOT adequate as backup
  • Confirm a portable oxygen cylinder is present, separate from the wall supply, with an attached flowmeter and cylinder wrench
  • Open the cylinder to confirm it is adequately filled; then close the valve

SECTION 2 - Electrical Power (D)

Item 2/3: Turn on Machine and Confirm AC Power
  • Confirm the machine is plugged into AC mains power
  • Turn on the anesthesia delivery system and all required electrical equipment
  • Some machines perform an automatic battery test during startup; verify battery backup function if not automated
  • Modern workstations run automatic self-check routines - confirm all automated checks pass

SECTION 3 - Suction (E)

Item: Verify Patient Suction
  • Confirm functioning suction with tubing of adequate length
  • Confirm a Yankauer suction tip is present and functional
  • Test suction strength before every case

SECTION 4 - Monitors & Alarms (E)

Item 4: Verify Monitors and Alarms
  • Confirm availability and function of all ASA standard monitors:
    • Pulse oximeter (with probe)
    • Capnograph (with sampling line connected)
    • Oxygen analyzer
    • Respiratory volume monitor (spirometer)
    • Breathing-system pressure monitor with high and low alarms
    • Blood pressure cuff (appropriate size)
    • ECG
  • Enable all alarms, set thresholds to departmental defaults, and confirm audible alarm volume is sufficient

SECTION 5 - Gas Supply: High-Pressure System (D)

Item 5: Spare O2 Cylinder on Machine
  • Open the O2 cylinder mounted on the back of the machine
  • Verify adequate pressure (≥ half full, ~1000 psig for a full E-cylinder)
  • Close the cylinder valve after checking
Item 6: Pipeline Gas Pressures
  • Confirm all pipeline hoses are connected to the correct wall outlets (check pin-index or DISS fittings)
  • Verify pipeline gauge pressures are ≥ 50 psig (typically ~50-55 psig)
  • If N2O or air is used, confirm those pipeline supplies are also connected and at adequate pressure

SECTION 6 - Low-Pressure System (D + E)

Item 7: Vaporizers (E)
  • Confirm vaporizer(s) are correctly seated and locked in place on the manifold
  • Check agent fill level; refill if needed
  • Tightly close all filler ports
  • Confirm only one vaporizer is enabled at a time
Item 8: Low-Pressure Leak Test (D)
  • Close all flow control valves and turn all vaporizers off
  • Attach the negative-pressure suction bulb to the common gas outlet
  • Squeeze the bulb until fully collapsed; it should remain collapsed for ≥ 10 seconds (no leak)
  • Repeat with each vaporizer turned on individually
  • Remove the bulb and reconnect the fresh gas hose
On machines WITHOUT an outlet check valve, a positive-pressure breathing circuit test will also detect low-pressure leaks.

SECTION 7 - Flow Meters & Hypoxic Guard (D)

Item 7: Test Flowmeters
  • Turn on the machine master switch
  • Attempt to dial a hypoxic O2/N2O mixture (< 21% O2)
  • Verify the anti-hypoxic device prevents this (flow changes or alarm activates)
  • Confirm all flowmeter bobbins/balls move freely and read zero at minimum flow

SECTION 8 - Scavenging System (D)

Item 8: Waste Gas Scavenging (WGSS)
  • Verify proper connections between the scavenging system and:
    • The APL (pop-off) valve
    • The ventilator relief valve
  • Adjust waste-gas vacuum to the recommended setting
  • Fully open the APL valve and occlude the Y-piece
  • With minimum O2 flow: allow scavenger reservoir bag to collapse - absorber pressure gauge should read ~0 cm H2O
  • With O2 flush active: allow scavenger bag to fully distend - absorber pressure gauge should read < 10 cm H2O

SECTION 9 - Breathing System (E)

Item 9: Calibrate O2 Monitor
  • Expose the O2 sensor to room air and confirm reading = 21%
  • Verify the low-O2 alarm is enabled and functioning
  • Reinstall the sensor in the circuit; flush with O2 and confirm reading > 90%
  • Calibrate or verify calibration of the capnometer and anesthetic agent analyzers
Item 10: Inspect Breathing Circuit
  • Set the selector switch to Bag mode
  • Confirm the breathing circuit is complete, undamaged, unobstructed, and properly assembled
  • Verify CO2 absorbent (soda lime/Amsorb) is not exhausted (not discolored, not hardened)
  • Install any accessory equipment to be used (humidifier, PEEP valve, HME filter)
Item 11: Breathing System Leak Test (E)
  • Set all gas flows to zero (or minimum)
  • Close the APL valve and occlude the Y-piece
  • Pressurize the circuit to ~30 cm H2O using the O2 flush
  • Confirm pressure remains stable for ≥ 10 seconds (no leak)
  • Open the APL valve and verify pressure drops promptly to zero
Item 12: APL Valve and Unidirectional Valve Function
  • Test APL valve: after pressurizing the circuit, open the APL valve and confirm rapid pressure release
  • Confirm unidirectional (inspiratory and expiratory) valves are seated and moving correctly

SECTION 10 - Ventilator (E)

Item 12/13: Test Manual and Automatic Ventilation
  • Place a test lung or second breathing bag on the Y-piece
  • Set appropriate ventilator parameters for the next patient
  • Switch to ventilator mode; turn on the ventilator
  • Fill the bellows and breathing bag with O2 flush; set O2 flow to minimum
  • Confirm the bellows fills and empties correctly during the ventilator cycle
  • Confirm the breathing bag on the Y-piece moves appropriately (ventilator tidal volumes visible)
  • Switch back to bag (manual) mode and confirm bellows ceases cycling
  • Test the O2 flush button - confirm no major change in tidal volume (no "flush dumping")
  • Perform the flow test: manually ventilate the test lung in a to-and-fro motion; confirm inspiratory valve opens during inspiration and expiratory valve opens during exhalation - feel for obstruction in either limb

SECTION 11 - Final Checks (E)

Item 13: Airway Equipment
  • Functioning laryngoscope with backup blade
  • Appropriate endotracheal tubes (with stylet) and oropharyngeal/nasopharyngeal airways
  • Magill forceps
  • Difficult airway supplies as appropriate (videolaryngoscope, LMAs, bougie)
Item 14: Medications & IV Supplies
  • Induction agents, opioids, muscle relaxants drawn up and labeled
  • Resuscitation drugs available (ephedrine, epinephrine, atropine, succinylcholine)
  • IV fluids and giving sets ready
Item 15: Special Equipment (D)
  • Defibrillator functional and immediately accessible
  • Difficult airway kit available
  • Malignant hyperthermia (MH) kit immediately accessible (if applicable)
  • Emergency drug kit confirmed

SECTION 12 - Documentation & Time-Out (E)

Item 14: Document Completion
  • Record completion of the checkout in the anesthetic record
  • Departments should maintain a log for quality assurance purposes
Item 15: Anesthesia Time-Out
  • Confirm ventilator settings are appropriate for the patient
  • Review the surgical safety checklist (Sign-In)
  • Confirm patient identity, procedure, allergies, NPO status, and site marking
  • Confirm team briefing is complete before induction

Quick Summary: The 15 Items of the 2008 ASA PAC

#ItemFrequency
1Backup ventilation & separate O2 sourceDaily
2Patient suction adequateEvery case
3Turn on machine, confirm AC powerDaily
4Monitors present and alarms functionalEvery case
5Spare O2 cylinder pressure adequateDaily
6Pipeline gas pressures ≥ 50 psigDaily
7Vaporizers filled, ports closedEvery case
8Low-pressure system leak testDaily / vaporizer change
9Scavenging system functionalDaily
10O2 monitor calibratedEvery case
11CO2 absorbent not exhaustedEvery case
12Breathing system leak testEvery case
13Gas flow through circuit (flow test)Every case
14Document checkoutEvery case
15Confirm ventilator settings & time-outEvery case

Sources:
This is a shared conversation. Sign in to Orris to start your own chat.