I now have comprehensive data from two leading anesthesia textbooks (Miller's Anesthesia 10e and Morgan & Mikhail's Clinical Anesthesiology 7e) plus the Canadian Anesthesiologists' Society 2022 guidelines. Here is the full compiled checklist:
Anesthesia Machine Pre-Use Checklist
Based on the 2008 ASA Recommendations for Pre-Anesthesia Checkout (PAC), as detailed in Miller's Anesthesia 10e and Morgan & Mikhail's Clinical Anesthesiology 7e.
Note: Because modern anesthesia workstations differ widely, no single procedural checklist applies to all machines. Each department should develop a machine-specific version based on these principles and the manufacturer's manual. Items marked * must be completed before every case, not just daily.
FREQUENCY KEY
- D = Daily (or when machine is moved / vaporizer changed)
- E = Before every case
SECTION 1 - Emergency Backup (D)
Item 1: Verify Backup Ventilation & Oxygen
- Confirm a self-inflating manual resuscitation bag (e.g., Ambu bag) is present and functional at every anesthetizing location
- A non-self-inflating Mapleson circuit is NOT adequate as backup
- Confirm a portable oxygen cylinder is present, separate from the wall supply, with an attached flowmeter and cylinder wrench
- Open the cylinder to confirm it is adequately filled; then close the valve
SECTION 2 - Electrical Power (D)
Item 2/3: Turn on Machine and Confirm AC Power
- Confirm the machine is plugged into AC mains power
- Turn on the anesthesia delivery system and all required electrical equipment
- Some machines perform an automatic battery test during startup; verify battery backup function if not automated
- Modern workstations run automatic self-check routines - confirm all automated checks pass
SECTION 3 - Suction (E)
Item: Verify Patient Suction
- Confirm functioning suction with tubing of adequate length
- Confirm a Yankauer suction tip is present and functional
- Test suction strength before every case
SECTION 4 - Monitors & Alarms (E)
Item 4: Verify Monitors and Alarms
- Confirm availability and function of all ASA standard monitors:
- Pulse oximeter (with probe)
- Capnograph (with sampling line connected)
- Oxygen analyzer
- Respiratory volume monitor (spirometer)
- Breathing-system pressure monitor with high and low alarms
- Blood pressure cuff (appropriate size)
- ECG
- Enable all alarms, set thresholds to departmental defaults, and confirm audible alarm volume is sufficient
SECTION 5 - Gas Supply: High-Pressure System (D)
Item 5: Spare O2 Cylinder on Machine
- Open the O2 cylinder mounted on the back of the machine
- Verify adequate pressure (≥ half full, ~1000 psig for a full E-cylinder)
- Close the cylinder valve after checking
Item 6: Pipeline Gas Pressures
- Confirm all pipeline hoses are connected to the correct wall outlets (check pin-index or DISS fittings)
- Verify pipeline gauge pressures are ≥ 50 psig (typically ~50-55 psig)
- If N2O or air is used, confirm those pipeline supplies are also connected and at adequate pressure
SECTION 6 - Low-Pressure System (D + E)
Item 7: Vaporizers (E)
- Confirm vaporizer(s) are correctly seated and locked in place on the manifold
- Check agent fill level; refill if needed
- Tightly close all filler ports
- Confirm only one vaporizer is enabled at a time
Item 8: Low-Pressure Leak Test (D)
- Close all flow control valves and turn all vaporizers off
- Attach the negative-pressure suction bulb to the common gas outlet
- Squeeze the bulb until fully collapsed; it should remain collapsed for ≥ 10 seconds (no leak)
- Repeat with each vaporizer turned on individually
- Remove the bulb and reconnect the fresh gas hose
On machines WITHOUT an outlet check valve, a positive-pressure breathing circuit test will also detect low-pressure leaks.
SECTION 7 - Flow Meters & Hypoxic Guard (D)
Item 7: Test Flowmeters
- Turn on the machine master switch
- Attempt to dial a hypoxic O2/N2O mixture (< 21% O2)
- Verify the anti-hypoxic device prevents this (flow changes or alarm activates)
- Confirm all flowmeter bobbins/balls move freely and read zero at minimum flow
SECTION 8 - Scavenging System (D)
Item 8: Waste Gas Scavenging (WGSS)
- Verify proper connections between the scavenging system and:
- The APL (pop-off) valve
- The ventilator relief valve
- Adjust waste-gas vacuum to the recommended setting
- Fully open the APL valve and occlude the Y-piece
- With minimum O2 flow: allow scavenger reservoir bag to collapse - absorber pressure gauge should read ~0 cm H2O
- With O2 flush active: allow scavenger bag to fully distend - absorber pressure gauge should read < 10 cm H2O
SECTION 9 - Breathing System (E)
Item 9: Calibrate O2 Monitor
- Expose the O2 sensor to room air and confirm reading = 21%
- Verify the low-O2 alarm is enabled and functioning
- Reinstall the sensor in the circuit; flush with O2 and confirm reading > 90%
- Calibrate or verify calibration of the capnometer and anesthetic agent analyzers
Item 10: Inspect Breathing Circuit
- Set the selector switch to Bag mode
- Confirm the breathing circuit is complete, undamaged, unobstructed, and properly assembled
- Verify CO2 absorbent (soda lime/Amsorb) is not exhausted (not discolored, not hardened)
- Install any accessory equipment to be used (humidifier, PEEP valve, HME filter)
Item 11: Breathing System Leak Test (E)
- Set all gas flows to zero (or minimum)
- Close the APL valve and occlude the Y-piece
- Pressurize the circuit to ~30 cm H2O using the O2 flush
- Confirm pressure remains stable for ≥ 10 seconds (no leak)
- Open the APL valve and verify pressure drops promptly to zero
Item 12: APL Valve and Unidirectional Valve Function
- Test APL valve: after pressurizing the circuit, open the APL valve and confirm rapid pressure release
- Confirm unidirectional (inspiratory and expiratory) valves are seated and moving correctly
SECTION 10 - Ventilator (E)
Item 12/13: Test Manual and Automatic Ventilation
- Place a test lung or second breathing bag on the Y-piece
- Set appropriate ventilator parameters for the next patient
- Switch to ventilator mode; turn on the ventilator
- Fill the bellows and breathing bag with O2 flush; set O2 flow to minimum
- Confirm the bellows fills and empties correctly during the ventilator cycle
- Confirm the breathing bag on the Y-piece moves appropriately (ventilator tidal volumes visible)
- Switch back to bag (manual) mode and confirm bellows ceases cycling
- Test the O2 flush button - confirm no major change in tidal volume (no "flush dumping")
- Perform the flow test: manually ventilate the test lung in a to-and-fro motion; confirm inspiratory valve opens during inspiration and expiratory valve opens during exhalation - feel for obstruction in either limb
SECTION 11 - Final Checks (E)
Item 13: Airway Equipment
- Functioning laryngoscope with backup blade
- Appropriate endotracheal tubes (with stylet) and oropharyngeal/nasopharyngeal airways
- Magill forceps
- Difficult airway supplies as appropriate (videolaryngoscope, LMAs, bougie)
Item 14: Medications & IV Supplies
- Induction agents, opioids, muscle relaxants drawn up and labeled
- Resuscitation drugs available (ephedrine, epinephrine, atropine, succinylcholine)
- IV fluids and giving sets ready
Item 15: Special Equipment (D)
- Defibrillator functional and immediately accessible
- Difficult airway kit available
- Malignant hyperthermia (MH) kit immediately accessible (if applicable)
- Emergency drug kit confirmed
SECTION 12 - Documentation & Time-Out (E)
Item 14: Document Completion
- Record completion of the checkout in the anesthetic record
- Departments should maintain a log for quality assurance purposes
Item 15: Anesthesia Time-Out
- Confirm ventilator settings are appropriate for the patient
- Review the surgical safety checklist (Sign-In)
- Confirm patient identity, procedure, allergies, NPO status, and site marking
- Confirm team briefing is complete before induction
Quick Summary: The 15 Items of the 2008 ASA PAC
| # | Item | Frequency |
|---|
| 1 | Backup ventilation & separate O2 source | Daily |
| 2 | Patient suction adequate | Every case |
| 3 | Turn on machine, confirm AC power | Daily |
| 4 | Monitors present and alarms functional | Every case |
| 5 | Spare O2 cylinder pressure adequate | Daily |
| 6 | Pipeline gas pressures ≥ 50 psig | Daily |
| 7 | Vaporizers filled, ports closed | Every case |
| 8 | Low-pressure system leak test | Daily / vaporizer change |
| 9 | Scavenging system functional | Daily |
| 10 | O2 monitor calibrated | Every case |
| 11 | CO2 absorbent not exhausted | Every case |
| 12 | Breathing system leak test | Every case |
| 13 | Gas flow through circuit (flow test) | Every case |
| 14 | Document checkout | Every case |
| 15 | Confirm ventilator settings & time-out | Every case |
Sources:
- Morgan & Mikhail's Clinical Anesthesiology, 7e, Table 4-3 (FDA Anesthesia Apparatus Checkout Recommendations)
- Miller's Anesthesia, 10e, Chapter 20 - Items 1-15 of the 2008 ASA Pre-Anesthesia Checkout Recommendations
- Canadian Anesthesiologists' Society Pre-Anesthetic Checklist (2022), based on 2008 ASA guidelines