Sop of cbct machine in veterinary

Searching the Web

CBCT machine standard operating procedure veterinary clinic

Searching the Web

cone beam CT veterinary SOP radiation safety protocol

Here is a comprehensive Standard Operating Procedure (SOP) for a CBCT Machine in a Veterinary Setting, compiled from veterinary radiology guidelines, ACVR standards, and clinical best practices:

Standard Operating Procedure (SOP)

Cone Beam Computed Tomography (CBCT) - Veterinary Practice

Document No.: VET-CBCT-SOP-001 Effective Date: [Insert Date] Review Due: [Insert Date - typically annual] Approved By: [Veterinarian-in-Charge / Radiation Safety Officer]

1. PURPOSE

This SOP establishes the procedure for the safe and effective operation of the CBCT machine in a veterinary facility, ensuring diagnostic image quality, patient and staff safety, and compliance with radiation protection regulations.

2. SCOPE

This SOP applies to all veterinary professionals, veterinary nurses/technicians, and students who operate or assist with CBCT scanning of animal patients (small animals, exotics, equids, etc.).

3. RESPONSIBILITIES

RoleResponsibility
Veterinarian-in-ChargePrescribe scans, interpret images, ensure clinical justification
Radiation Safety Officer (RSO)Oversee compliance, maintain dosimetry records
Veterinary Technician / NurseOperate the CBCT unit, position patients, monitor anaesthesia
All StaffFollow ALARA principles, wear PPE, report faults

4. PREREQUISITES AND PREREQUISITES CHECKLIST

Before any CBCT procedure:
  • Valid veterinary prescription / clinical indication documented
  • Patient identity confirmed (species, name, weight, patient ID)
  • Anaesthesia/sedation plan confirmed with supervising vet
  • Relevant medical history reviewed (prior imaging, allergies, contraindications)
  • Radiation dosimetry badge (TLD/OSL badge) worn by all staff
  • PPE available: lead aprons, thyroid shields, lead gloves if needed

5. PERSONAL PROTECTIVE EQUIPMENT (PPE) REQUIREMENTS

  1. All personnel in the scanning room during scout/positioning must wear a lead apron (min. 0.25-0.5 mm Pb equivalent) and thyroid collar.
  2. Dosimetry badges must be worn at collar level (outside apron) at all times.
  3. During X-ray emission, all staff must exit the room or stand behind the lead-lined barrier.
  4. Line-of-sight anaesthesia monitoring must be maintained (visual/camera/monitoring equipment).
  5. PPE must be inspected at least annually for cracks or leaks and replaced as needed.

6. MACHINE SETUP AND PRE-SCAN CHECKS

6.1 Daily Startup

  1. Power on the CBCT unit per manufacturer instructions (e.g., NewTom 7G VET, VetCAT, Planmeca ProMax).
  2. Allow the unit to complete its self-diagnostic warm-up cycle.
  3. Log in to the imaging workstation and confirm software (NNT, CS 3D, or equivalent) is operational.
  4. Verify detector calibration is current; run flat-field/dark-field calibration if prompted.
  5. Check that the patient table/stretcher is clean and free of artefact-causing objects.
  6. Confirm the room is clear of non-essential personnel.

6.2 Quality Control (QC)

  • Perform daily QC scan using phantom (if available) per manufacturer protocol.
  • Log all QC results in the Equipment QC Register.
  • Report any image quality degradation, artefacts, or mechanical issues to the RSO immediately - do not use the machine until cleared.

7. PATIENT PREPARATION

  1. Anaesthesia/Sedation: All veterinary CBCT scans require the patient to be under general anaesthesia or deep sedation to eliminate motion artefact.
    • Confirm anaesthesia induction is complete and airway is secured before positioning.
    • Attach monitoring equipment (pulse oximeter, capnograph, ECG).
  2. Positioning: Select patient position based on anatomical region of interest:
    • Dental/maxillofacial: Dorsal recumbency (sternal) with head in gantry, mandible parallel to the detector plane.
    • Skull/ears: Lateral or dorsal recumbency per diagnostic need.
    • Body/limbs (for units with body CBCT capability): Position on stretcher according to manufacturer's anatomical reference guide.
    • Use radiolucent foam pads and positioning aids - no metallic objects near the scan field.
    • Remove all collars, leads, metallic devices, and dental metalwork if possible.
  3. Field of View (FOV) Selection:
    • Use the smallest FOV that covers the area of interest to minimise radiation dose (ALARA principle).
    • For dental: typically 8x8 cm or 10x10 cm FOV.
    • For full skull: 17x17 cm or larger (machine-dependent).
    • Document selected FOV in patient record.
  4. Scan Protocol Selection:
    • Select appropriate kVp and mA based on patient size and weight.
    • Use automatic exposure control (SafeBeam™ or equivalent) when available.
    • Choose Low Dose (ECO) protocol where diagnostic quality is still achieved.
    • Document all scan parameters (kVp, mA, FOV, scan time, exposure time).

8. SCANNING PROCEDURE

  1. Position the patient and confirm gantry clearance - no part of the patient should touch the gantry during rotation.
  2. Perform a scout (localiser) image to confirm correct positioning before full scan.
  3. Ensure all personnel have exited the scan room before initiating X-ray emission.
  4. Confirm the radiation warning light/buzzer is functioning outside the room.
  5. Initiate the scan from the control console.
  6. Monitor the patient via anaesthetic monitoring equipment (camera or window view) throughout the scan (scan time: typically 7-26 seconds).
  7. On scan completion, re-enter the room promptly and attend to the patient.
  8. Transfer image data to workstation and confirm image acquisition was successful before moving the patient.

9. POST-SCAN PROCEDURE

  1. Recover the patient from anaesthesia per facility anaesthesia recovery SOP.
  2. Label and save the DICOM dataset with correct patient ID, species, date, and referring vet details.
  3. Create the 3D reconstruction using appropriate software (NNT viewer, OnDemand3D, Dolphin 3D, or equivalent).
  4. Archive images per the facility's data management policy (minimum retention as per local regulations - typically 7 years).
  5. Generate a diagnostic report or refer images to a board-certified veterinary radiologist/dentist as appropriate.

10. RADIATION SAFETY - ALARA PRINCIPLES

Following ACVR radiation safety guidelines and the ALARA (As Low As Reasonably Achievable) framework:
PrincipleAction
Minimise TimeKeep exposure time as short as clinically acceptable; use lowest dose protocol
Maximise DistanceAll staff leave the room during emission; radiation falls off exponentially with distance
ShieldingUse lead-lined room barriers; wear lead aprons and thyroid shields
CollimationUse smallest FOV appropriate for the clinical question
JustificationOnly scan when clinical benefit outweighs radiation risk
  • Dosimetry badge results must be reviewed monthly by the RSO.
  • Cumulative dose records must be maintained for all radiation workers.
  • Pregnant staff must inform the RSO immediately and be temporarily reassigned from radiation duties.

11. INFECTION CONTROL

  1. Clean and disinfect the patient table, head support, and all contact surfaces between patients with an appropriate veterinary disinfectant.
  2. Use disposable covers on positioning aids where possible.
  3. Follow standard handwashing/hand hygiene protocols before and after patient handling.

12. FAULT REPORTING AND EMERGENCY PROCEDURES

  • If the machine malfunctions mid-scan, abort the scan via the emergency stop and attend to the patient first.
  • Do not attempt repairs - contact the manufacturer's service engineer.
  • Log all faults in the Equipment Fault Register with date, description, and action taken.
  • If a radiation incident occurs (e.g., accidental beam exposure to staff), report immediately to the RSO and follow the facility's radiation incident reporting procedure.

13. DOCUMENTATION AND RECORDS

RecordFrequencyResponsible
Patient scan record (kVp, mA, FOV, dose)Per scanTechnician
QC logDailyTechnician
Dosimetry badge recordMonthlyRSO
PPE inspection logAnnualRSO
Equipment maintenance logPer serviceRSO / Engineer
Staff radiation training recordAnnualPractice Manager

14. TRAINING REQUIREMENTS

  • All operators must complete documented training on the specific CBCT unit before independent use.
  • Annual radiation safety refresher training is mandatory for all staff working with ionising radiation.
  • Training records must be retained and made available for regulatory inspection.

15. REFERENCES


Note: This SOP must be reviewed annually or whenever the equipment, software, or relevant regulations change. Adapt specific parameters (kVp, mA ranges, FOV sizes) to your machine model and manufacturer specifications. Consult your local radiation authority for licensing and regulatory compliance requirements specific to your country.

Sop standard for cbct with newtom 5xl machine series wise

Searching the Web

NewTom 5XL CBCT machine specifications scan protocols FOV series SOP

Searching the Web

NewTom 5XL operator manual kVp mA scan parameters technical specifications

This is a shared conversation. Sign in to Orris to start your own chat.